PROGRAMMING…
7.6
Set Up Thermocouple Page
Set Up Thermocouple
Page header
Thermocouple Type
Set the thermocouple type required: Type
K
,
R
,
S
or
NONE
. If thermocouple
type is set to
NONE
, a preset temperature must be used – see below.
Preset Temperature
If a thermocouple is not used, set the process temperature.
[600 to 1400
°
C in 1
°
C steps]
Note. This frame is not displayed unless
Thermocouple Type
is set to
NONE
.
Press
to return to the top of the
Oxygen Calibration Page.
Press
to advance to the
Set Up Outputs Page.
Preset Temp
700
-----
SET UP T/C
-----
T/C Type NONE
K
SET UP OUTPUTS
R
S