TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 11
Reset to Factory Configuration
This function resets all of the device parameters to the factory set-
tings. Refer to the product instruction for information on the factory
settings.
Key
Display
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select
YES
.
Select
RESET CAL
.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
E N T E R
CALIBRATION
PROCESS CAL
EDIT CAL
→
RESET CAL
E N T E R
WORKING
CALIBRATION
COMPLETE
ID TAGNAME
READY