TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 7
CALIBRATE PROCESS TEMPERATURE
This section steps through calibrating the process temperature.
Calibration complete.
Key
Display
Comments
CALIBRATION
COMPLETE
E N T E R
ID TAGNAME
READY
Key
Display
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select
YES
.
Select
PROCESS CAL
.
Select
TEMPERATURE
.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
E N T E R
CALIBRATION
→
PROCESS CAL
EDIT CAL
RESET CAL
E N T E R
PROCESS CAL
CONDUCTIVITY
→
TEMPERATURE