98
| NG C820 0 AN D PGC1000 US ER MAN UA L | 2101510MNA H
The visual examination should also verify there is no damage to the cast aluminum explosion-
proof enclosure.
10.3 Technical support and returns
If technical assistance is required during maintenance or for returning parts, contact ABB customer
service. See the contact information at the end of this document.
If an ABB component is to be returned for repair, securely wrap it in protective anti-static packaging.
Before returning a component, call for a return authorization number (RA). Affix this number to the
outside of the return package. Use the RMA form (see the link provided in the
section).
Parts shipments must be prepaid by the customer. Any part not covered by original system warranty
will be shipped to the customer F.O.B.
NOTICE – Equipment damage.
The end caps have precision-engineered threading and
are susceptible to damage if treated roughly. Take care when removing housing covers.
Explosion-proof threads are coated with thread lubricant. Reapply appropriate thread
lubricant as needed.
10.4 Spare part components
The information in this section presents the components and parts that are accessible for removal and
installation.
10.4.1
Replacement components
The following components may be replaced (see
:
Analytical module (12 or 24 VDC) with or without GC module (see
)
GC module
Digital controller assembly with display
Termination panel
Feed-through assembly without preheat (see
)
Feed-through assembly with preheat (12 or 24 VDC)
10.4.2
Replacement parts
The following parts may be replaced:
Lithium Battery
Frit Filters
Analytical processor to termination panel cable
Termination panel to digital controller cable
Feed-through O-ring
Feed-through interface gasket
Feed-through manifold gasket
Feed-through heater (12 or 24 VDC)
GC module temperature sensor