2101510MNA H | NGC82 00 AN D P GC 1000 US ER MA NUA L |
55
Figure 6-17: PCCU setup for local Ethernet communication
7.
Click
Close
to close connection setup.
8.
Click
Entry
on the PCCU32 menu bar to connect to the device.
9.
Click the
Entry
icon on the PCCU32 toolbar to connect to the device. If unable to connect, see
section
8.6 Basic Ethernet connection troubleshooting
Figure 6-18: PCCU32
10. Click
Yes
if the message to synchronize the date and time displays. This message displays if
the analyzer’s calendar clock does not match the laptop's date and time which is usually the
case with a new system.
When connection is successful, the analyzer’s startup wizard screen displays. The wizard provides the
sequence to complete basic analyzer configuration. Go to section
6.4
Power on sequence
6.4.1
Initialization
At power on, the analyzer starts its initialization sequence:
Performs a cold-start
Loads startup information into RAM memory. The message “Boot Loader” displays.
Start Diagnostics. Diagnostic results are available on the Diagnostic tab in PCCU. Once the
initializing sequence is complete, connect with the analyzer using PCCU.
6.4.2
Startup diagnostics
The analyzer has an extensive built-in list of tests which are automatically performed each time the
device is started.