Navigator 600 Silica
Single-stream
5 Configuration
30
IM/NAV6S/SS–EN Rev. E
5.3.5 User
Fields
Description
If 'Security system' is set to 'Basic' this tab lists the four users, User 1 to User 4. Selecting a user opens a new page with two
fields:
Name – the user’s name, up to 20 characters.
Password – each user can be assigned a unique 4-digit security code for Configuration level access.
If 'Security System' is set to 'Advanced' and User 1 (administrator) is logged on, the 'User' tab shows additional fields:
User 1 Name
User 1 identification tag – up to 20 characters.
User 1 Access
A page opens with two checkboxes to select whether User 1 has 'Calibration and Maintenance' and
/ or 'Logging access'.
User 1 Password
User 1's password – a unique 20-character (alphanumeric) security code. A minimum password
length applies.
View / Edit Other Users
Selects the other user's access levels and passwords. If selected additional fields appear:
User X Name
Where X is the user number (2 to 12) – up to 20 characters can be used.
User X Access
Where X is the user number (2 to 12). A dialog box is displayed listing the access available for the
user:
Calibration & Maintenance
Logging
Configuration (No access)
Configuration (Load)
Configuration (Limited)
Configuration (Full)
User X Password
Where X is the user number (2 to 12). The password for User X.
If 'Security System' is set to 'Advanced' and a user other than User 1 is logged on, the 'User' tab has three fields.
These fields can be edited only if User 1 has set the security field 'Reconfigure preset' to 'Yes' – See page 29.
Where X is the user number (2 to 12).
User X Name
User X identification tag. Up to 20 characters.
User X Access
Where X is the user number (2 to 12). A dialog box is displayed listing the access available for the
user:
Calibration & Maintenance
Logging
User X Password
User X's password – a unique 20-character (alphanumeric) security code. A minimum password
length applies.