Navigator 600 Silica
Single-stream
4 Installation
IM/NAV6S/SS–EN Rev. E
15
4.4 Reagent Bottle Location
4.4.1 Behind Mounting Plate or Wall
4.4.2 To Side of Analyzer
To prevent reagent flow due to gravity during maintenance,
ensure top of reagent bottles are below the level of the pumps.
Ensure reagent tubes reach the bottom of all 4 reagent bottles.
4.4.3 Below Analyzer
Mount the reagent bottles below the front of the analyzer and
install a protective cover (not supplied) over the reagent bottles.
Caution. Do not
place reagent bottles on the floor.
Dimensions in m (ft.)
Fig. 4.2 Reagent Bottles Mounted Behind
the Mounting Plate or Wall
Ma
xi
mum 1 (3.28
)
Reagent tube length
1.3 (4.26)
Dimensions in m (ft.)
Fig. 4.3 Reagent Bottles Mounted to the Side of the Analyzer
Dimensions in m (ft.)
Fig. 4.4 Reagent Bottles Mounted Directly Below the Analyzer
Maximum 1 (3.28)
Rea
g
ent tube
length 1.3 (
4
.26)
Ensure top of reagent bottles
are below level of pumps
Max
imum 1 (3.28)
R
e
agent tube le
ngth 1.3
(4.26)
Protective Cover
(not supplied)