Navigator 600 Phosphate
Multi-stream
5 Configuration
IM/NAV6P/MS–EN Rev. F
39
5.8.2 email 1 and email 2
The analyzer can be configured to send emails to a maximum of 6 recipients in response to certain events. The addressees can all
subscribe to the same SMTP server or the analyzer can be configured to send emails via 2 different SMTP servers to a maximum of 3
addressees per server.
Up to 10 independently configurable triggers may be enabled to generate an email when the selected source becomes active. When
a trigger source becomes active, an internal 1 minute delay timer is started. At the end of that minute, an email is generated that
includes, not only the event that initiated the delay timer, but every other event that occurred during the delay period together with any
enabled reports. The data returned in the email therefore reflects the real-time alarm state at the time the email was generated, not
the state when the first trigger source became active.
Each email sent includes a link to the analyzer's embedded web server, enabling the analyzer's data and status to be viewed remotely
using an internet browser on a PC.
Fields
Description
SMTP Server IP Address
The IP address of the SMTP server through which emails are to be routed.
Recipient 1 to Recipient 3
The email address of recipient 1 to 3.
Inverted Triggers
The option to invert triggers 1 to 6.
Trigger 1 to 10
The trigger for an email message to be sent. A dialog box is displayed with three options:
None
– no triggers are set.
Archive state
– a dialog box is displayed with six options:
–
Archive media not present
–
Too many files on the archive media
–
Archive media 100% full
–
Archive media 80% full
–
Archive media present
–
Archive on-line
Event group
– a dialog box is displayed with nine options:
–
Solutions (reagent, cleaning solution, secondary solution is low or out)
–
Samples (out of one or more samples)
–
Optics (faulty optics)
–
Temperatures (control block / electronics out of limits)
–
Power (excessive current)
–
Calibration (factor / offset error or missed calibration)
–
Hardware (temperature sensor, analog-to-digital converter 1 to 3 failures)
–
Service (service due / media card full)
–
Alarm Relays (alarm relay active)