Navigator 600 Phosphate
Multi-stream
5 Configuration
IM/NAV6P/MS–EN Rev. F
27
5.3.4 Security
'User 1' is the
System Administrator
and is the only user that has access to the 'Security type' parameter. User 1 details cannot be
changed by any other user.
There are two types of security:
Basic
Enables up to four users to have access to the 'Configuration' menu. Each user has a password that can have up to four digits.
A separate password can be set to gain access to the 'Calibration & Maintenance' and 'Logging' menus. All four users use this
password.
Advanced
Enables up to twelve users to have password-protected access to any of the 'Configuration', 'Calibration & Maintenance' or
'Logging' menus. Each user can be assigned a unique 20-digit (alphanumeric) case-sensitive password. A minimum password
length can be set.
User 1 can set permissions for each user to access the 'Calibration & Maintenance' and / or 'Logging' menus. Access to the
'Configuration' menu can also be set – see Section 5.3.5, page 29.
–
Passwords are set by User 1. If permissions are set, users can subsequently change their own passwords.
–
Password expiry dates can be set.
–
Inactive user accounts can be disabled after a set time.
–
Password failure limits can be set.
–
A minimum password length can be set.
Note.
A separate password for access to the 'Calibration & Maintenance' and 'Logging' menus is not required.