3 Calibration
3.11. Calibration, axis 4
57
3HAC 022907-001 Revision: -
3.11. Calibration, axis 4
General
This section details how to perform the actual fine calibration of axis 4 using the alternative
method with Levelmeter 2000.
Required equipment
Equipment
Art. no.
Info
Levelmeter 2000
6369 901-347
(incl. one
sensor)
Only one sensor required.
Sensor plate
3HAC 0392-1
Turning disk fixture
6808 0011-GU
Includes guide pins.
Sync adapter
3HAB 7981-1
Required for robots:
IRB 2400/10
IRB 2400/16
Sync adapter
3HAB 1525-1
Required for robot:
IRB 4400
Sync adapter
3HAC13908-1
Required for robots:
IRB 6600
IRB 6650
IRB 6650S
IRB 7600
Sync adapter
6896 134-GZ
Required for robot:
IRB 6400R
Guide pin
-
Used for guiding the sync adapter to the
turning disk.
Dimensions:
•
IRB 4400: D=8, L=16
•
IRB 6600, IRB 6650, IRB
6650S, IRB 7600: D=12, L=20
•
IRB 6400R: D=10, L=20
Isopropanol
1177 1012-108
For cleaning the attachment points.
Other tools and procedures may
be required. See references to
these procedures in the step-by-
step instructions below.
These procedures include references to
the tools required.
Summary of Contents for Levelmeter 2000
Page 2: ......
Page 6: ...Table of Contents 3HAC 022907 001 Revision 4...
Page 8: ...Overview 3HAC 022907 001 Revision 6...
Page 14: ...1 Overview 1 4 Calibration movement directions for all axes 3HAC 022907 001 Revision 12...
Page 78: ...3 Calibration 3 17 Updating revolution counters 3HAC 022907 001 Revision 76...
Page 82: ......