Parameterization
D184B141U02
FSM4000
EN - 37
6.2
Entering data in “short form”
Task
Action
Display
information
→
V98.14 %
Starting point for “Process information”
→
V12.000 m
3
Any parameter
1.
Open menu.
C/CE
is displayed
"Prog. level"
2.
Search "Prog. Level".
ENTER
Disabled
"Prog. level"
3.
Switch off “Programming protection”
Select "Specialist".
STEP or DATA
Specialist
"Prog. level"
4.
Set value.
ENTER
Specialist
Direct numeric entry
Entry from table
Task
Input
Display information
Task
Input
Display
information
Qmax
Submenu
5.
Find parameter
"Qmax".
STEP or
DATA
1800.00
m³/h
5.
Parameter “Submenu
unit”.
STEP or
DATA
Unit
Qmax
Unit Qmax
6.
Change parameter
"Qmax".
ENTER
m³/h
6.
Change parameter
m³/h.
ENTER
m³/h
Min 500 m³/h
Max 10000 m³/h
Unit Qmax
7.
Find desired value,
e.g., m³/min.
STEP or
DATA
m³/min
Qmax
6 2 4 0 , 0 0 m³/h
8.
Set value.
ENTER
Unit Qmax
m³/min
Unit Qmax
9.
A new value is
displayed.
m³/min
10.
Exit menu.
C/CE
7.
Enter the desired
value in the sequence.
6 x DATA
STEP
2 x DATA
STEP
4 x DATA
STEP
STEP
10 x DATA
STEP
STEP
8.
Set value.
ENTER
Qmax
6240.00
m³/h
9.
A new value is
displayed.
Qmax
6240.00
m³/h
C/CE
10.
Exit menu.
"Prog. level"
11.
Search "Prog. Level".
STEP or DATA
Specialist
"Prog. level"
12.
Switch on “Programming protection”.
ENTER
Specialist
"Prog. level"
13.
Select "Locked"
C/CE
Locked