Parameterization
OI/FEX300/FEX500-EN FEX300,
FEX500
123
Menu / Parameter
Value range
Description
Diagnostics /
Trend
1)
Conductivity
Electrode Coat. QE1
Electrode Coat. QE2
The measured values are shown as a line diagram.
When the "Trend-Logging" function is enabled, the
measured values are stored at the defined interval
(cycle time). The last 12 measured values are stored
and indicated as a line diagram. The oldest data
record is overwritten.
Logtime Interval
1 ... 45,000 min.
Interval for measured value logging.
Trend-Logging
ON / OFF
Activate the "Trend-Logging" function.
When the "Trend-Logging" function is enabled, the
measured values are stored at the defined interval
(cycle time).
The data records can be read out by the
"ScanMaster" verification software and analyzed as
trends.
Diagnostics /
Flowrate Alarm
Max. Flowrate Alarm
0 ... 130 %
Set the maximum limit value for the flow.
Min. Flowrate Alarm
0 ... 130 %
Set the minimum limit value for the flow.
Diagnostics /
Output Readings
Current mA
DO1 Frequency
Hz
DO2 State
open / closed
Digital In State
open / closed
Display the current values and statuses of the listed
inputs and outputs.
italics
= Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.