Endura APA592
pH / Redox (ORP) transmitter
7 Operator Pages and Menus
OI/APA592–EN Rev. E
59
2-Point Manual Calibration
Step
Procedure
Screen
1
At the
PV Manual Cal
parameter select
the
2-Point Manual
calibration option –
this screen provides an option to hold the
output.
Press the
key to cancel hold or
the
key to hold the output.
2
Edit the buffer temperature by pressing
the
key and use the
and
keys
to select the required temperature.
Press the
key to proceed
.
3
Edit the 1
st
buffer value by pressing
the
key and use the
and
keys
to select the required value.
Press the
key to proceed
.
4
Apply Buffer 1 input, watch for PV stability
(wait for the message '
Continue When
Stable
' to be displayed) then press
the
key to continue.
5
Edit the 2
nd
buffer value by pressing
the
key and use the
and
keys
to select the required value.
Press the
key to proceed
.
6
Apply Buffer 2 input, watch for PV stability
(wait for the message '
Continue When
Stable
' to be displayed) then press
the
key to continue.
Table 7.3 2-Point Manual Calibration