Endura APA592
pH / Redox (ORP) transmitter
7 Operator Pages and Menus
58
OI/APA592–EN Rev. E
1-Point Manual Calibration
Step
Procedure
Screen
1
At the
PV Manual Cal
parameter select
the
1 Point Manual
calibration option and
apply an input.
Wait for the message '
Continue When
Stable
' to be displayed then press
the
key to continue.
2
Enter the new PV value using the
and
keys then press the
key to
continue.
3
At the calibration results screen(s) check
the new values.
Press the
key to exit
the screen:
or
if a calibration failure message is
displayed, reset the PV value (see steps 1
and 2).
Table 7.2 1-Point Manual Calibration