Endura APA592
pH / Redox (ORP) transmitter
4 Electrical Installation
30
OI/APA592–EN Rev. E
4.4 Sensor Cable Color-Coded Connections
4.4.1 Standard Sensors with No Diagnostic Functions
Note.
ORP (Redox) and Antimony pH sensors do not have temperature compensation and therefore do
not have temperature sensors or related wiring.
Note.
Turn
Off
diagnostics when connecting these sensors
Connection Information
pH Sensor Cable Color
Transmitter
Terminal
Function
2867
AP100
AP300
7650/
60*
TB5
4
Glass / Metal Sensing Electrode
Clear
Clear
Blue
Clear
Blue
5
Guard
–
–
–
–
–
6
Reference
Black
Black
Black
Black
Black
7
Solution Ground
–
–
–
–
–
8
Temperature Compensator
(If included with sensor)
–
Red
Red
Red
Red
9
Temperature Compensator
(If included with sensor)
–
White
White
White
White
10
Shield /Shield (if fitted)
–
–
–
Yellow
–
11
Temperature Compensator
(If included with sensor)
–
Red
Grey
Red
–
*Cut back the 7650 / 60 sensor's redundant green wire.