Chapter 5
Set Parameters
Fundamentals
General Procedure
The normal procedure for setting parameters is:
1. Press
Enter
to access the menu.
2. Select the Parameters menu item with the
×
or
Ø
key.
3. Activate the Parameters menu item with the
Enter
key.
4. Change the parameters with the
×
or
Ø
key.
5. Confirm the change with the
Enter
key or cancel the change with the
Back
key.
6. Use the
Back
key to return to the previous menu.
Only parameters that blink in the display can be edited.
The following example should clarify the details of this general procedure.
Step Action
Key Display
Key
CO 105.2 ppm
1
Enter
Calibrate
Access the menu.
2
×
Ø
Setting parameters
Enter
Select the “Setting
parameters” menu.
3
×
Ø
Parameters component
CO
Enter
Select the submenu.
4
×
Ø
Span setpoint
505.0 ppm CO
Enter
Select the parameter to
be changed.
5
×
Ø
Span setpoint
505.0
ppm CO
Enter
Edit the parameter value
and confirm the change.
6
Span setpoint
500.0 ppm CO
Back
Return to the previous
menu.
7
Parameters component
CO
Back
Return to the previous
menu.
8
Setting parameters
Back
Return to the measured
value display.
CO 105.2 ppm