EL6010 Gas Analyzer in Category 2G Operator’s Manual
42/24-30 EN Rev. 1
Gas Inlet and Outlet Conditions,
continued
Sample Gas
Dew Point
The sample gas dew point should be at least 5 °C below the lowest ambient
temperature throughout the sample gas path. Otherwise a sample gas cooler or
condensate trap is required. Water vapor content variations cause volume errors.
Corrosive Gases
The EL6010 gas analyzers must not be used for the measurement of corrosive
gases. Gases like chlorine (Cl
2
) and hydrogen chloride (e.g. wet HCl) as well as
gases or aerosols containing chlorine must be cooled or undergo prior absorption.
Gas Outlet Conditions
The outlet pressure must not exceed the inlet pressure by more than the pressure
loss across the analyzer.
Items Delivered
Quantity Description
1
EL6010-CU Control Unit
1
EL6010-... Analyzer Unit with attached connection cables for
24 VDC power supply and data transmission
1
Analyzer Data Sheet
1
Operator’s Manual 42/24-30 EN
1
2.5 mm Hex Wrench
Items Delivered
1
220 mm diam. x 3 mm O-Ring
Analyzer Data Sheet
Contents of the
Analyzer Data Sheet
The analyzer data sheet contains the following information:
•
Order number (A-No.)
•
Part number (P-No.)
•
Serial number (F-No.)
•
Production date
•
Power supply (voltage, frequency, consumption)
•
Software version
•
Sample components and measuring ranges
•
Serial numbers of installed components
Keep the Analyzer
Data Sheet Safely
The analyzer data sheet must be kept on the gas analyzer so that it is always
ready-to-hand.