CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
14
Decommissioning
Decommissioning the gas analyzer
In the case of a temporary shutdown:
1. Shut off the sample gas.
2. Purge the gas lines and gas feed paths in the gas analyzer
with dry air or nitrogen for at least 5 minutes.
3. Switch off the gas analyzer power supply.
In the case of a long-term shutdown, carry out the following in
addition:
4. Remove the gas lines from the gas analyzer ports. Tightly
seal the gas ports.
5. Disconnect the electrical leads from the gas analyzer.
Fidas24 – Shutting down the gas analyzer
In the case of a temporary shutdown:
1. Shut off the sample gas supply at the sampling point.
2. Purge the sample gas line with nitrogen for at least
5 minutes, from the sampling point.
3. Set the gas analyzer to standby mode, refer to
In case of corrosive and combustion gases set the gas
analyzer to standby operation with detector purging.
4. Shut off combustion air supply and combustion gas
supply.
In the case of a long-term shutdown, carry out the following in
addition:
5. Shut off instrument air supply.
6. Switch off the gas analyzer power supply.
7. Remove the gas lines from the gas analyzer ports. Tightly
seal the gas ports.
8. Disconnect the electrical leads from the gas analyzer.