GAA610-M
ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
91
Figure 57: Electrical connections SCC-F
8. When the cover is removed and the power is disconnected,
you can remove the marked connectors
1
. Pay attention to
the PE on the connection plate. All connections to the SCC-F
are now disconnected.
9. Remove the four screws that connect the SCC-F to the
GAA610-M.
10. The SCC-F can now be removed.
Prepare the new SCC-F
Figure 58: Transport lock SCC-F (Bottom view)
1. Unscrew the mounting brackets from the rear of the side
covers and screw them securely at the front of the side
covers, flush with the front cover, using the drill holes
provided for this purpose.
2. Use a Ph2 crosshead screwdriver to loosen the transport
lock
1
(two M6×25 mm screws) in the base plate.
3. Unscrew all 8 screws of the cover (Figure 56
). Remove the
cover and connection plate. Pay attention to the PE
connections.
Figure 59: Required connections SCC-F, electronics number
4. Remove the marked connectors
1
and make sure that the
230 V setting
2
is selected.
5. Make a note of your individual electronics number
3
.
6. The SCC-F can now be installed.