Aztec ATS430 | Turbidity sensor
|
OI/ATS430–EN Rev. H 7
4 Sensor setup – first-time installation
To perform a first-time installation (
Easy Setup
menu):
1.
Connect a new or replacement sensor to the transmitter’s
EZLink connector – see transmitter Operating instructions
OI/AWT440-EN.
The following prompt is displayed identifying the new /
replacement sensor (S1 to S4):
2. To enter
Easy Setup
level, press the
key
(below the
icon).
The
Easy Setup
start screen is displayed:
Press the
key (below the
Select
prompt).
3. To enter
Easy Setup
level, press the
key
(below the
Select
icon).
4. Press the
key (below the
Edit
prompt) to change the
default value to the required value / selection.
5. Press the
key (below the
Next
prompt) to accept the
value / selection displayed and advance to the next
configuration parameter.
The following
Configuration
parameters are set at
Easy
Setup
level:
— Tag
— PV Type
— Units
— Range High
— Range Low
— Clean Interval
— Filter Type
— Analogue outputs and alarms
6. Continue with configuration of the required parameters.
On completion the
Easy Setup
start screen is displayed:
7.
To exit
Easy Setup
, press the
key (below the
Exit
prompt) to display the
Operator Page
.
Pressing the
key (below the
Select
prompt) re-enters the
Easy Setup
level where parameters can be reviewed or
modified after 1st time connection.
After completing the
Easy Setup
level, pressing the
or
key enters the
Advanced Configuration
level, where
all available sensor and transmitter parameters can be
reviewed or modified.
IMPORTANT (NOTE)
Perform this procedure when a new / replacement
sensor is connected to the transmitter for the first time
only. For existing sensors, see Section 5, page 8.
ATS400
New Sensor Sx
Detected
Start Easy
Setup ?
00-00-2014
Menu
Easy Setup
Exit
Select
IMPORTANT (NOTE)
Refer to Section 5.1, page 9, for parameter details – not
all parameters in Section 5.1 are displayed at
Easy Setup
level.
IMPORTANT (NOTE)
To re-configure an existing sensor (after first-time
installation), enter the
Configuration
level (see
Section 5.1, page 9) via the
Operator Page
– refer to
transmitter Operating instructions
Operator Page
details and navigation.
Menu
Easy Setup
Exit
Select