52
AW T4 20 |
U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER | O I/AW T4 2 0 - EN R E V. B
…14
Sensor setup
4-electrode conductivity
Menu
Comment
Default
Tag
Enter an alphanumeric sensor tag (16 characters maximum) to identify the sensor on the
Operator Pages
.
TAG1
Measurement Type
Select measurement type:
• Conductivity/Concentration
Note
. If a change is made the I/O sources are reset.
Conductivity
Note .
The following menus are displayed only if
Measurement Type
=
Conductivity
Conductivity Unit
Select the conductivity units:
• mS/cm/μS/cm
mS/cm
Sensor Group
Enter the sensor group for the measuring cell used –
• Group A/Group B
see the relevant conductivity cell manual.
Group A
Range High
Set the span value used in
Chart
and
Bargraph
views.
Sensor Group dependent
– see table below
Range Low
Set the zero value used in
Chart
and
Bargraph
views.
0
Note .
The following menus are displayed only if
Measurement Type
=
Concentration
Sensor Group
Enter the sensor group for the measuring cell used –
• Group A/Group Bw
see the relevant conductivity cell manual.
N/A
Conc . Solution
Note. Displayed only if
Sensor Group
=
Group A
Select the Concentration Solution
• NaOH/HCl/H2SO4/H3PO4/NaCl/KOH/Custom
N/A
Concentration Unit
Note
. Displayed only if
Conc. Solution
=
Custom
Select the Concentration Units
• None(Blank)/ppm/mg/l/ppb/μg/l/%/Custom
N/A
Custom Units
Note
. Displayed only if
Concentration Units
=
Custom
Enter an alphanumeric string (6 characters maximum) for the custom (user defined) concentration units.
N/A
Conc . Curve Table
Set the user defined concentration curve using the 6-point linearizer table
(concentration against conductivity).
N/A
Range High
View the span value used in
Chart
and
Bargraph
views.
N/A
Range Low
View the zero value used in
Chart
and
Bargraph
views.
N/A
Filter Type
Select the signal filtering type:
• None/Low/Medium/High
None
Temp . Comp . Type
Set the type of temperature compensation:
• Manual/Automatic/None
Automatic
Manual Temperature
Note
. Displayed only if
Temp. Comp. Type
=
Manual
Enter the temperature of the sample within the range –10.0 to 120.0 °C.
25.0 °C
TC Curve
Note
. Not displayed only if
Temp. Comp. Type
=
None
Set the type of automatic temperature compensation required:
• TC Coeff./Standard KCl/NaOH/NaCl/HCl/H2SO4/H3PO4/KOH/User Defined
TC Coeff
User Def . TC Curve
Note
. Displayed only if
TC Curve
=
User Defined.
Set the user defined temperature compensation curve using the six point linearizer table (% against °C).
N/A
TC Coefficient
Note
. Displayed only if
TC Curve
=
User Defined.
Enter the temperature coefficient (α x 100) of the solution (0.01 to 5.00 %/°C).
If unknown, the temperature coefficient (α) of the solution must be calculated – see page 97.
2.00 %/°C
Sensor Diagnostics
Dirty Sensor
Detect dirty sensor condition:
• Enabled/Disabled
Disabled
Out Of Solution
Detect Out Of Solution condition:
Enabled/Disabled
Disabled
Reset To Defaults
Select to reset all
Sensor Setup
parameters to their default values
Sensor group
Conductivity measuring range
A
0 to 2000 mS/cm
B
0 to 2000 µS/cm