Settings
3-9
Configure for Static Address Assignment
To use static address assignment for the HM5, your local network administrator must
supply the necessary information.
1.
Change the DHCP setting to
Disabled
. (You can ensure
Disabled
is set by
selecting the setting value.)
The data entry fields in the New Settings area then become available.
2.
Enter the following settings using the values from your network administrator:
IP Address
Subnet Mask
Gateway
DNS
3.
Press
Accept
.
3.6.4
Host Network
When configured for Host Network communication, the HM5 can send results to another computer on
the network (unidirectional mode), or accept work orders from and send results to a computer
(bidirectional mode). The computer must be running compatible software to accept the result
information and optionally send order information. For a list of compatible software, visit the Abaxis
website at
.
3.6.4.1
Personal Computer (PC)
The
Host Network > Personal Computer (PC)
settings enable the HM5 to send results one-way
(unidirectional) to a network-enabled Personal Computer (PC) system or software such as a practice
management system. In this mode, the HM5 cannot accept work list orders (bidirectional
communication) from a PC system.
When sending results, the HM5 establishes an outbound connection to a
specified IP address and port (see below). The receiving PC must be
running a service that listens for incoming connections on the IP address
and port. The local network administrator or software vendor can provide
the necessary IP address and port values.
To configure the HM5 to send results to a PC system:
1.
In the Host Network screen, set the Networked PC field to
Active
.
2.
Enter the PC Remote IP Address and PC Remote Port number, as
provided by your local network administrator or software vendor.
Summary of Contents for VetScan HM5
Page 1: ...Operator s Manual ...
Page 2: ......
Page 4: ...1 2 ...
Page 6: ...1 4 ...
Page 26: ...2 10 Installing the VetScan HM5 ...
Page 38: ...3 12 Settings ...
Page 50: ...4 12 Running a Sample ...
Page 64: ...5 14 Quality Control and Calibration ...
Page 80: ...6 16 Maintenance Service ...
Page 88: ...8 4 Shutdown ...
Page 112: ...9 24 Troubleshooting ...
Page 126: ...A 10 Introduction to Veterinary Hematology ...
Page 136: ...B 10 Operating Principles ...
Page 154: ...D 16 Veterinary Case Studies ...
Page 163: ......