Installing the VetScan HM5
2-9
2.3 Initializing the VetScan HM5
1.
Select
Measure > Run
. A Message that the blank has expired will appear.
2.
Select
OK
to run a blank.
The analyzer will run a blank measurement. When the process is complete, the instrument dis-
plays the result of the blank measurement. (Blank runs are described in more detail in
ning a Blank Measurement” on page 4-5
.)
3.
Select
Accept
if there are no flags and no values highlighted in red, indicating that the analyzer
blank background level is within specifications.
The analyzer then displays a sample measurement screen, as shown, and
is now ready to per-
form an analysis
.
4.
Verify the analyzer’s system settings, and make any needed changes: see
Note:
You will be notified with a pop-up if the blank
falls outside specifications. If this occurs,
refer to
problem persists, call Abaxis Technical Sup-
port.
Summary of Contents for VetScan HM5
Page 1: ...Operator s Manual ...
Page 2: ......
Page 4: ...1 2 ...
Page 6: ...1 4 ...
Page 26: ...2 10 Installing the VetScan HM5 ...
Page 38: ...3 12 Settings ...
Page 50: ...4 12 Running a Sample ...
Page 64: ...5 14 Quality Control and Calibration ...
Page 80: ...6 16 Maintenance Service ...
Page 88: ...8 4 Shutdown ...
Page 112: ...9 24 Troubleshooting ...
Page 126: ...A 10 Introduction to Veterinary Hematology ...
Page 136: ...B 10 Operating Principles ...
Page 154: ...D 16 Veterinary Case Studies ...
Page 163: ......