2-6
Installing the VetScan HM5
19.
Select a
Pack Setup
:
If the reagent pack will be
on the
same level
as the HM5 (preferred) and
you have the QCA (Quick Change
Apparatus), select
Quick Change
(QCA)
in
Reagent Replace.
If the reagent pack will be
on a lower level
than the
HM5 and/or the Bottle Caps tubing set is being
used, select
Bottle Caps
. Make sure the top of the
reagent pack is no more than 18 in (45.7 cm) lower
than the bottom of the analyzer.
20.
Select
Schedule Settings > Reset > Accept
to restart maintenance timers from today.
21.
Pour ~200 ml of distilled water into the Abaxis flask.
22.
Submerge the free ends of the reagent tubes into the
water.
23.
Select
Maintenance > Reagent Status
, and then
Prime All
.
24.
Run
Prime All
twice more (three times in all) until all
water is gone from the flask and air is pumped into
the tubing.
25.
While the analyzer is priming, open the reagent pack.
Remove the bottle caps from each of the five bottles. Save these lids for later disposal of the
pack.
Summary of Contents for VetScan HM5
Page 1: ...Operator s Manual ...
Page 2: ......
Page 4: ...1 2 ...
Page 6: ...1 4 ...
Page 26: ...2 10 Installing the VetScan HM5 ...
Page 38: ...3 12 Settings ...
Page 50: ...4 12 Running a Sample ...
Page 64: ...5 14 Quality Control and Calibration ...
Page 80: ...6 16 Maintenance Service ...
Page 88: ...8 4 Shutdown ...
Page 112: ...9 24 Troubleshooting ...
Page 126: ...A 10 Introduction to Veterinary Hematology ...
Page 136: ...B 10 Operating Principles ...
Page 154: ...D 16 Veterinary Case Studies ...
Page 163: ......