9-14
Troubleshooting
9.2.8
MicroDilutor Motor Error
A MicroDilutor Motor error indicates an issue with the motor that draws blood into the sample needle.
Correct this error as follows:
1.
Turn the analyzer off and open its side door.
2.
Locate the cylinder (circled), place a finger on it and spin it back and
forth several times.
3.
Turn the analyzer back on and attempt to run a blank.
4.
If the blank runs, perform an Soak Cleaning.
9.2.9
Diluent and Lyse Dilutor Motor Errors
These errors refer to the motors that move the syringes in the back of the analyzer.
Correct this error as follows:
1.
Check for kinks in the reagent tubing from the analyzer to the bottle caps or dip tube heads.
2.
If no kinking is found, turn the analyzer off, open the back door of
the analyzer and manually move the syringes up and down a few
times.
If they move, turn the analyzer back on and attempt to run a
blank.
If they do not move or appear to be damaged, call Abaxis
Technical Support.