ABACUS MPA2A Instructions For Use Download Page 1

Parameter Setting Example via Menu 

To Adjust Hi alarm value 

 

Press 

mute

 & 

adj

 together and hold 

 
 

Press 

res 

(once) 

 
 

Release 

mute

 & 

adj 

 

 

Press 

hi

 7 times 

(display will read HI ALM = (Value)) 

 
 

Press adj 

 
 

Press adj 

(display will read ADJUST THE VALUE) 

 
 

Press 

hi

 to increase value or 

lo

 to decrease value 

 
 

Press 

set

 

(within30 seconds or display will go back to normal & changes will be ignored) 

 
 
 
 
 

For Quick Set of hi and lo Alarm Setpoint  

 

Press hi or lo 

 

Press adj 

Use hi or lo to adjust value 

Press adj to select ‘ENABLED’ 

 

Press set 

(within 30 seconds or display will go back to normal & changes will be ignored) 

 

 
 
 
 
Abacus Instruments

13C Old Bridge Way, Shefford, Beds.,  England ,

 

SG17 5HQ 

Phone; +44 (0)1462 638200  Fax; +44 (0)14612 638201 www.abacusinstruments

.

com

 
MPA2A Alarm

 

Instructions for use 
 

 

 

Note:  

The internal battery link between terminals 6 & 7 has been removed for shipping. 

This link must be remade for the unit to operate. 

 
Connections:- 

 
240V ac Supply 

Terminal 1 (Live) 

 

 

Terminal 2 (Neutral) 

 

 

Terminal 3 (Earth) 

 
Battery Link 

  Terminal 6 & 7 

 
Alarm Relay 

Terminal 8 (Common) 

 

 

Terminal 9 (N/C) 

 

 

Terminal 10 (N/O) 

 

 

 Note: The relay is fail safe hence N/C contacts 

 

 

are closed when unit is in alarm condition 

 
Buzzer  

Terminal  11 + 

 

 

 Terminal  12 – 

 
Defrost Switch 

  Terminal  13 & 14 (Note Closed cct = Defrost) 

 
Door Switch 

 Terminal  15 & 14 (Note Closed cct = Alarm) (Factory selectable) 

 
PT1000 sensor 

Terminal 23 Red (GND) 

 

 

Terminal 24 White 

 

     