4
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
27 of 138
Operating Instructions — Hardware
Start Up the System
Before the system is turned on, make sure the site requirements specified in the
Site Planning
Guide
are met. This guide includes information on the site layout, electrical, gas supply,
ventilation and waste collection, and environmental requirements.
1. Make sure that there is clear access to the AC mains supply cable. The AC mains
supply cable must be accessible in order to disconnect the mass spectrometer from
the AC mains supply.
2. Make sure that the 4 L source exhaust drain bottle is connected to the Exhaust
Waste connection on the rear of the mass spectrometer and to the laboratory
ventilation system.
3. Make sure that the AC mains supply cable is plugged in to the mass spectrometer.
4. Make sure that the mass spectrometer and roughing pump AC mains supply cables
are plugged into the AC mains supply.
5. Make sure that the Ethernet cable is connected to both the mass spectrometer and
the computer.
6. Turn on the roughing pump. The mains switch is located beside the mains supply
input connection on the roughing pump.
7. Wait 5 minutes, then turn on the mass spectrometer switch.
8. Turn on the computer, if it was turned off.
9. Start the Analyst
®
software.
Reset the System
•
Press and hold the
Reset
button for 5 seconds.
An audible click is heard when the relay activates. After approximately three
minutes, the mass spectrometer should reach operating pressure.
WARNING! Lifting Hazard: Risk of personal injury or system damage. If the
system must be moved, contact an FSE.
Note:
Before operating the system, read the safety information in the
Precautions and Limitations on page 9
.
Note:
If the switch is already on, turn it off and then on.
Summary of Contents for QTRAP 4500 LC/MS/MS system
Page 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Page 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Page 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Page 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...