Page 82
July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
3.6.7
Column 3
Figure 3-35: Column 3
C3 Enable Column
C3 Enable Surface Reference
C3 Cell Size
C3 Distance First Cell Center
C3 Number OF Cells
: Available from 1 to 25 cells. The instrument can measure multiple cells along the column from
instrument to the maximum range given by the scatter condition, or from surface and down to the instrument dependent
on the configuration.
C3 Cell Center Spacing
3.6.8
Column 4
Figure 3-36: Column 4
Enable Current From Wave Pings.
A profiler are normally not able to measure Acoustic Wave and Current
simultaneously because the sensor using the same transducer for both operations. However with the DCPS you might
set the
Enable Current From Wave Ping
which make it possible to use the data from Wave measurement pings to also
calculate a reduced column with current. This is presented in
Column 4
when enabled. This column is limited to only
measure in 3 different cells and these cells are controlled by the setting under
User Maintenance
and
Wave
Measurement,