July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
1.7.1 Factory Configuration
All properties in this section are Read Only, not possible to overwrite for the user. Only certified Aanderaa service
personal can alter these settings. The access level for reading the status of this properties are however different for
each property, see table for more details. In this group we find information about Software and hardware settings,
Production, Service and Calibration dates and limits for quality parameters where status will be shown in
Cell State 1,
Cell State 2
and
Record Status
. See chapter 5.6.2 for more details.
Table 1-5: Sensor properties for Doppler Current Profiler Sensor 5400/5400R, 5402/5402R, and 5403/5403R
ENUM=Enumeration, INT =Integer, BOOL=Boolean (‘yes’/’no’)
Property
Type
No of
e
lem
ent
s
Use
Conf
igur
ation
Ca
tegor
y
Acces
s
Pro
te
cti
on
R
ea
d/W
rite
Product Name
String 31 AADI Product name, default Doppler Current Profiler Sensor
FC
No /
Read
Only
Product Number String 6
AADI Product number, default 5400, 5400R, 5402, 5402R, 5300 or
5300R
Serial Number
INT
1
Serial Number
SW ID
String 11 Unique identifier for internal firmware
SW Version
INT
3
Software version (Major, Minor, Built)
HW ID X
String 19 Hardware Identifier, X =1..3
1)
unique identifier for hardware
High /
Read
Only
HW Version X
String 9
Hardware Identifier, X =1..3
1)
(Rev. x)
System Control INT
3
For AADI service personnel only
Production Date String 31 AADI production date, format YYYY-MM-DD (Not in use)
Last Service
String 31 Last service date, format YYYY-MM-DD, empty by default
Last Calibration String 31 Last calibration date, format YYYY-MM-DD (Not in use)
Calibration
Interval
INT
1
Recommended calibration interval in days (Not in use)
Cross Diff Limits INT
3
Set limits 1, 2, and 3 for “Cell has High cross difference, above
cross difference limit”, default values are 10, 25 and 40. If value
exceeds the limits a bit in Cell State1 will be set. See Cell State
Strength Limits
NB
INT
4
Set limits 1, 2, 3 and Discard Data
2)
for “Cell has weak signal,
strength below weak limit for narrowband”, default values are -40, -
44, -48 and -50. If value exceeds the limits a bit in Cell State1 will
be set. If Beam Strength Output is set a status for each individual
beam will be shown in Cell State 2. See Cell State definition in