12.4. Restoring the factory set values
If the gravity acceleration value or calibration weight value is changed unintentionally,
restore those values to the factory set values, as follows.
1. Refer to “12.1. Calibration mode” to enter
the calibration mode. The gravity
acceleration value is displayed.
2. Press the UNITS key twice to display
“
ClrC
”
.
.
3. Press the TARE key to display
“
ClrC
no
” with “
no
” blinking.
4. Press the PRINT key.
“
ClrC no
” changes to “
ClrC go
” with
“
go
” blinking.
To cancel the restoring procedure, press the
ZERO key. The display returns to step 2.
5. When “
ClrC go
” is displayed, press the
TARE key. The factory set values are
restored and “
end
” is displayed.
To finish the setting procedure, press the
ZERO
1
key or the CAL switch. The scale
returns to the weighing mode.