14
Leveling foot
Spirit level
The size and shape
depends on the scale model.
Display knobs
Display
Display
Display knobs
7.2. Installing the scale
1. Select the place for installing the scale.
Refer to “4. PRECAUTIONS”.
2. Adjust the level of the base, using the
spirit level and leveling feet. Confirm
that the bubble is in the center of the
level.
SW type with a display pole has an
extra foot at the bottom of the pole.
Adjust this foot to touch the floor after
adjusting the level of the base.
3. Adjust the display angle as follows.
Type with a display pole
(1) Rotate the two display knobs
located on the display rear to
remove them.
(2) Tilt the display to an appropriate
angle, 0°, 30°, 60° or 90°.
(3) Attach the two display knobs
and secure them firmly.
Type with a display stand
(1) Loosen the two display knobs
located on the sides of the
display.
(2) Tilt the display to an appropriate
angle. (Free setting angle)
(3) Tighten the two display knobs.
4. Confirm that nothing is touching the
keys and connect the AC power cable
to an electrical outlet.
DANGER
•
The base frame is connected to the power GND (EXC-) inside the load
cell. Be sure to earth ground the AC power cable. When the AC electrical
outlet has no ground terminal, use the ground terminal (screw with the
ground symbol) on the scale.
Location of the ground terminal
Models with a display pole: lower part of the display pole
Models with no display pole: bottom of the weighing pan