6.Measuring Input Signals - 6.3.Starting and Ending Recording
66
1WMPD4004444
6.3.2. Starting and Ending Recording
Starting Recording
Press the START key on the operation panel to start
device recording and enclose the screen with a red
frame.
When the memory waveform is displayed on the monitor,
you can display the Trigger tab from CH on the
side menu to display [Memory block] for checking the
state of recording to the memory block.
Each time a trigger is detected and recording to the
memory block is performed, the display is updated with
the latest waveform and used block count.
Ending Recording
Recording ends when the recording time set in
Recording time elapses or when the STOP key on
the operation panel is pressed. Then the monitor screen
automatically switches to the [PLAYBACK] screen and
displays the last recording data. The screen is enclosed
in a green frame on the [PLAYBACK] screen.
When the memory waveform is displayed on the monitor,
you can display the Trigger tab from CH on the
side menu to display [Memory block] for checking the
recorded memory data.
You can tap Display block in the memory waveform
playback to change the recorded block with the rotary
knob.
Tap and change the display block with the rotary knob
Red frame
Green frame