![A&D GC Series Instruction Manual Download Page 36](http://html.mh-extra.com/html/aandd/gc-series/gc-series_instruction-manual_446421036.webp)
36
13.3.
Correcting Gravity Acceleration
When the scale is first used or has been moved to a different place, sensitivity adjustment needs to be
performed using a weight. But if the weight for sensitivity adjustment cannot be arranged, change the gravity
acceleration of your scale to the value of the area where the scale is used. The gravity acceleration setting is
not required when sensitivity adjustment using a weight is performed.
1. The scale displays “
” and the gravity acceleration value setting.
2. Use the numeric keys 0 to 9 to enter the gravity acceleration value.
To correct the value, press the C key and re-enter it.
(Refer to “
20.3. Gravity Acceleration Map
3. Press the ENTER key.
This stores the gravity acceleration that has been input and
returns to the “
” display.
4. Turn on the display again by pressing the ON/OFF key
or the RESET key.
13.4.
Initializing Sensitivity Adjustment Value
When sensitivity adjustment is changed due to incorrect operation, the settings for gravity acceleration and
sensitivity adjustment using a weight should be returned to the factory settings.
1. The scale displays “
”.
2. Every time the PRINT key is pressed, “ ” and “ ” are
shown alternately.
3. Press the ENTER key with “ ” displayed to return the
sensitivity adjustment value to the factory setting. After “
”
is shown, the scale enters the weighing mode.
Press the ENTER key with “ ” displayed to show
“
” again.
Gravity acceleration value