51
11-2 GLP Report
Set the following parameters to output the report.
To print the report, set the “GLP output (
info
)” parameter to “
1
”, set the “Data output pause
(
pU5e
)” parameter to “
1
” and use MODE 3 of the AD-8121B. For details on using the printer,
refer to “15-1 Connection to the AD-8121B Printer”.
To output the report to a personal computer using the RS-232C serial interface, set the “GLP
output (
info
)” parameter to “
2
”.
Note
If the time and date are not correct, adjust the AD-8121B clock and calendar (FX-
i
series) or
use “
Cl adj
“ of the function table to adjust the time and date (FZ-
i
series).
Calibration report using an internal weight
When the setting is “
info 1
”:
When the setting is “
info 2
”:
A & D< TE RM >
M ODE L FZ -300 0i <T ER M>
S /N 01 2345 67 <T ER M>
I D A BCDE FG <T ER M>
D ATE <T ER M>
20 08 /1 2/ 31<T ERM>
T IME <T ER M>
12 :3 4: 56<T ERM>
C ALI BR AT ED (INT .)<TER M>
S IGN AT UR E< TERM >
< TER M>
< TER M>
- - - - - - - - - - - - - - - - < TERM >
< TER M>
< TER M>
CALIB RATED( INT.)
SIGNA TURE
- - - - - -
A & D
MODEL
S/N
ID
DATE
TIME
FZ- 3000i
012 34567
AB CDEFG
2008/ 12/31
12: 34:56
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
AD-8121 format
General data format