32
8-6 Calibration Test Using an External Weight
This function tests the balance weighing accuracy using an external weight and outputs the result.
This is available only when the “GLP output (
info
)” parameter is set to “
1
” or “
2
”. (Calibration test
does not perform calibration.)
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until
CCout
is
displayed, and then release the key.
3 The balance displays
CC 0
.
z
If you want to change the target weight (a list of
usable weights is shown o n page 26), press the
SAMPLE key and go to step 4.
z
If you use the target weight value stored in the
balance, go to step 5.
4 Specify the target weight value as follows:
SAMPLE
key To switch the display
condition to: “All of the
segments blinking” (target
weight selection mode) or
“The last two digits blinking”
(value adjustment mode).
RE-ZERO key (To increase the value)
MODE key (To decrease the value)
To select the target weight or
adjust the value.
key
To store the new weight
value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.
CAL key
To cancel the operation and
return to
CC 0
.
Press and
hold the key
Release
the key
e.g. Target
weight value
1000.12 g
To next page
Select a
weight