47
Sttaannddaarrd Teem
mppeerraattuurree
Sttaannddaarrd PPrreessssuurree
Noorrm
maal Teem
mppeerraattuurree
Noorrm
maal PPrreessssuurree
Standard Temperature and Normal Temperature menu selections allow the
user to first select desired temperature units of measure: °C, °F, °K, or °R. By
default, currently active temperature units will be selected. Once the units are
selected, the use can adjust the desired temperature value and press the
joystick equivalent of an
ENT
button to save it. By default, currently stored
values will be displayed.
Standard Pressure and Normal Pressure menu selections allow the user to first select the
desired pressure units of measure. By default, currently active pressure units will be selected.
Once the units are selected, the user can adjust the desired pressure value and then press the
joystick equivalent of an
ENT
button to save it. By default, currently stored values will be
displayed.
6..4
4..113
3..2 D
Diis
sp
plla
ay a
an
nd P
Prro
oc
ce
es
ss IIn
nffo
orrm
ma
attiio
on ((P
PII)) S
Sc
crre
ee
en
ns
The local OLED Process Information screens can be configured to be static
(manual control) or dynamic (automatic sequencing). In the static mode,
pressing the joystick
UP
allows the user to page through the PI screens in the
forward direction, while pressing the joystick DN pages through the PI screens
in the opposite direction. When the last PI screen is reached, the firmware
“wraps around” and scrolls to the initial PI screen once again.
NOTE:
PI screens which are not Enabled (masked) will be skipped. PI
Screen #1 (Mass Flow Rate, Pressure, Temperature and Totalizer #1
reading) cannot be Disabled.
NOTE:
Once Standard Temperature/Pressure and/or Normal Temperature/
Pressure values are changed, the corresponding PI mass flow readings
shown on the instrument display or transmitted via digital or analog interface
will change as well.
NOTE:
Thefactory defaultvaluefor the Standard Temperatureis70
Û)
(21.1
Û&
, andforNormalTemperatureis32
Û)
(0.0
Û&
NOTE:
The factory default value for both Standard and Normal
Pressure is 14.696 PSIA (1.0 atm).