A.R.C. Laser C-Pro User Manual Download Page 31

Manual

C-PRO

31 

8  Technical Data 

8.1  Dimensions 

Fig. 18: Dimension of the C-Pro 

Dimensions: 260 mm x105 mm x 65 mm (length x width x height) 

Weight: 570 g 

8.2  General 

Designation 

Micromanipulator 

Type 

C-Pro

Transmission 

70 % ± 5 % 

Wavelength CO

2

 laser 

10600 nm 

Maximum power CO

2

 laser 

30W / CW, 10W / CW, Super Puls 

Operating distance 

200 - 500 mm 

Operating area 

Minimum 

Maximum 

50 x 30 mm at a 

working distance of 200 mm 

145 x 75 mm at a  

working distance of 500 mm 

Summary of Contents for C-Pro

Page 1: ...User Manual C Pro Micromanipulator Medical Device A R C LASER GmbH Bessemerstr 14 90411 N rnberg Germany phone 49 911 21779 0 fax 49 911 21779 99 info arclaser de...

Page 2: ...t A R C LASER GmbH Bessemerstr 14 D 90411 N rnberg Germany TEL 49 911 21779 0 FAX 49 911 21779 99 Revision Date Nov 22 2017 File Manual_C PRO_10600nm_rev3 docx Revisions Date 12 10 2020 Author Christi...

Page 3: ...or 12 4 4 Connecting the C Pro to the Surgical Microscope 12 4 4 1 Microscope compatibility 12 4 4 2 Connecting the C Pro 13 4 5 Connecting the Micromanipulator to the Laser 14 4 5 1 Connecting the ar...

Page 4: ...trolling the laser beam 29 7 4 Dismantling 29 7 5 Quick guide 30 7 6 C Pro performance test after installation 30 8 TECHNICAL DATA 31 8 1 Dimensions 31 8 2 General 31 9 SERVICE 32 9 1 Introduction 32...

Page 5: ...ntire user manual without an explicit written consent by A R C Laser GmbH Approved copy or parts extracted from this user manual as licensed by A R C Laser GmbH shall include references to the origina...

Page 6: ...laser beam The main fields of application are indications in the field of ENT and gynecology CAUTION The device may only be operated by specially trained personnel who have knowledge of the medical e...

Page 7: ...the beam in the field of vision Fig 1 Beam path at the micromanipulator 2 2 Mirror and Lens Based Systems Micromanipulators are divided into mirror based and lens based systems depending on their stru...

Page 8: ...e C Pro instead of lenses as is common with many other manufacturers If the operating field is very weakly illuminated two target beams may be seen due to the light refraction at the deflecting mirror...

Page 9: ...or the connection and the application of the micromanipulator see figure 4 Note The C Pro and all accessories should always be stowed in the appropriate recesses in the case after use The joystick and...

Page 10: ...ease inform A R C Laser GmbH as soon as possible in case parts are missing or damaged 3 3 Storage Conditions Transport and storage conditions Temperature 10 C to 60 C Relative humidity 5 85 The C Pro...

Page 11: ...o light up when the laser is in operation This ensures warning to prevent accidental entry to the room without protective goggles Please store the laser protective goggles at the entrance of the room...

Page 12: ...t the top and the underside of the deflection mirror for burn in and contamination Hold the micromanipulator in a firm grip 4 Note Never touch the coated side of the deflection mirror 5 If cleaning is...

Page 13: ...see figure 5 3 Push The dovetail adapter attached to the micromanipulator into the adapter plate so that the lens of the microscope is located centrally above the housing of the micromanipulator As a...

Page 14: ...e thread adapter Make sure the articulated arm is in the correct position see figure 7 8 The articulated arm is taken out of the holder by means of slight pressure of the two fingers so that the conne...

Page 15: ...trated operating area The laser must be positioned in such a way that the articulated arm is within the operating range shown in figure 10 4 6 2 Angular position of the articulated arm The two arms of...

Page 16: ...line with the angular range mark red green red label on the first leg see figure 11 right After the laser has been placed in the positions described in 4 5 1 and 4 5 2 the laser and the operating mic...

Page 17: ...r In addition you can refer to the regulations of the American National Standard Agency ANSI Z136 3 1996 American National Standard for the Safe Use of Lasers in Health Care Facilities and ANSI Z136 1...

Page 18: ...protected from catching fire This can be achieved by the use of fire resistant instruments venturi ventilation humidification and non burning gas mixtures The anesthesia staff should use fire resistan...

Page 19: ...address reorders to A R C Laser GmbH indicating the wavelength or the item number of the goggle Should you require any additional information on protective goggles please contact your local A R C Las...

Page 20: ...materials as this may cause a fire The pedal switch controlling the laser Pulse should never be outside the range of the operator holding the laser hand piece fiber It is prohibited that any person ot...

Page 21: ...company operates worldwide and considers the protection of the environment and natural resources as a corporate obligation Based on individual tests we can for the products of ARC Laser GmbH confirm t...

Page 22: ...been tested for operation in heights exceeding 2 000 m above sea level The following ambient conditions must be met Operating conditions Temperature 15 C to 35 C Relative humidity 75 not condensing A...

Page 23: ...n training all trained persons are mentioned in the device book wherein a person as responsible for the security is registered This person is subsequently entitled to instruct other people on the unit...

Page 24: ...s equipped with the following components see figure 13 Fig 13 Components on the C Pro connection for articulated arm adapter positioning ring operating distance dovetail focusing defocussing lever def...

Page 25: ...na Vario 1 x ME02157 thread adapter for articulated arm 1 x ME12033 hand rest long 1 x VP12000 transportation case C Pro 1 x ZU01084 cleaning set selected when ordering For information on accessories...

Page 26: ...the beam diameter is set the lower the energy density for the same laser parameters The power density is defined by the following formula with The unit of is In the table below the energy densities a...

Page 27: ...cused before the target beam can focused see figure 14 3 Get the laser ready Press the grey button standby in the lower right corner of the display It will change into yellow button ready 4 The desire...

Page 28: ...this lever position for tissue cutting Moving the lever in the direction D defocused will cause an increase in the beam diameter A defocused beam is used for coagulation and vaporization of tissue In...

Page 29: ...ted by turning the cap Fig 17 C Pro from the surgeon s perspective left focus defocus center joystick right hand rest The size of the operating area depends on the operating distance Minimum 50 x 30 m...

Page 30: ...be adjusted as required 7 6 C Pro performance test after installation To ensure that the C Pro is working well please perform the following test routine after installation 1 Put a wooden spatula in f...

Page 31: ...dth x height Weight 570 g 8 2 General Designation Micromanipulator Type C Pro Transmission 70 5 Wavelength CO2 laser 10600 nm Maximum power CO2 laser 30W CW 10W CW Super Puls Operating distance 200 50...

Page 32: ...trained by A R C Laser GmbH This includes also any type of optics cleaning within the laser system This includes also any type of optics cleaning within the laser system 9 2 Care and Maintenance by th...

Page 33: ...ning set ZU01084 Pre cleaning For cleaning the mirror please turn the micromanipulator over so that the underside of the deflecting mirror points upwards Fig 20 Position of the C Pro at the beginning...

Page 34: ...eflecting mirror by means of lens cleaning tissue 2 The cleaning tissue should overlap on the side facing the observer so that the observer can hold the tissue with the tweezers without touching the m...

Page 35: ...he lens cleaning tissue may only be used once a fresh tissue should be used for each additional cleaning operation The steps described should be repeated as often as necessary until the mirror surface...

Page 36: ...ination of the operating area Explanation see chapter 7 1 No target beam Switch on the laser Activate the target beam Check the operating distance of the adjusting ring Micromanipulator is loose Check...

Page 37: ...sing of the device shall be passed on to A R C Laser GmbH within 24 hours Parts returned during the warranty period upon the explicit request of A R C Laser GmbH are subject to the written confirmatio...

Page 38: ...Manual C PRO 38 Notes...

Page 39: ...Manual C PRO 39 Notes...