6.4 Spot Diameters and Energy Densities
The following table contains a list of the diameters of the focused laser beam as a function
of the operating distance
Table 3: Spot diameter depending on operating distance
Operating distance
Spot diameter focused
Spot diameter defocused
200 mm
100 µm
2.0 mm
250 mm
110 µm
2.1 mm
300 mm
120 µm
2.2 mm
350 mm
130 µm
2.5 mm
400 mm
140 µm
2.9 mm
450 mm
150 µm
3.4 mm
500 mm
160 µm
3.9 mm
Since spot size and laser power are set independently of each other, it is necessary to
check the power setting after changing the spot size as the energy density changes.
The energy density is the distribution of the power to a specific area of the treatment area.
The larger the beam diameter is set, the lower the energy density for the same laser
parameters.
The power density is defined by the following formula:
with
,
,
The unit of
is
In the table below, the energy densities are calculated for a selection of spot sizes and
capacities.
Table 4: power densities dependent on laser power and spot size
Laser
Power
P [W]
CO
2
Beam Spot Size d [mm]
0.1
0.2
0.3
0.4
0.5
1.0
2.0
3.0
4.0
1
12,732
3,183
1,415
796
509
127
32
14
8
2
25,465
6,366
2,829
1,592
1,019
255
64
28
16
3
38,197
9,549
4,244
2,829
1,528
382
95
42
24
4
20,929
12,732
5,659
3,183
2,037
509
127
57
32
5
63,662
15,915
7,074
3,979
2,546
637
159
71
40
10
127,324 31,831
14,147
7,958
5,093
1,273
318
141
80
15
190,985 47,746
21,221
11,937
7,639
1,910
477
212
119
20
254,647 63,662
28,294
15,915
10,186
2,546
637
283
159
25
318,309 79,577
35,368
19,894
12,732
3,183
796
354
200
30
381,971 95,493
42,441
23,873
15,279
3,820
955
424
239