background image

A.H. Systems Model Active Monopole Antennas 

 

 

 

 

©

 A.H. Systems inc

 

REV E 

4

 

 

SAFETY PRECAUTIONS

 

 
This equipment is intended for general laboratory use in a wide variety of industrial and 
scientific applications and designed to be used in the process of generating, controlling 
and measuring high levels of electromagnetic Radio Frequency (RF) energy. It is the 
responsibility of the user to assure that the device is operated in a location which will 
control the radiated energy such that it will not cause injury and will not violate regulatory 
levels of electromagnetic interference. To reduce the risk of fire, electric shock and/or 
injury to persons, basic safety precautions should always be followed when using 
electrical devices, including the following: 
 

1. 

READ ALL INSTRUCTIONS BEFORE OPERATING THE ACTIVE MONOPOLE 
ANTENNA.

 

 

2. 

Use the antenna only as specified.

 

 

3. 

Please dispose of batteries responsibly and in accordance with all applicable laws 
and regulations. 

 

4. 

To protect against electric shock, do not immerse the power cord, power plug or 
Antenna in water or in any other liquid.

 

 

5. 

Do not operate the product with a damaged charger cord or charger plug or after 
the antenna malfunctions or has been damaged in any manner.

 

 

6. 

This is not intended for permanent outdoor use.

 

 

7. 

Do not allow the cord or the antenna to rest on or near hot surfaces such as a hot 
gas or electric burner or a heated oven.

 

 

8. 

Disconnect the charger from the active monopole when not in use and before user 
servicing and cleaning.

 

 

9. 

Do not use this product for anything other than its intended use. 

 
 

RANGE OF ENVIRONMENTAL CONDITIONS 

 
This equipment is designed to be safe under the following environmental conditions: 

 
Indoor use 

 

Altitude: up to 2 km 

 

Temperature:  5

 C to 40

 C 

 

Maximum relative humidity: 80% for temperatures up to 31

 C. 

Decreasing linearly to 50% at 40

 C 

 

Pollution degree 2: Normally non-conductive with occasional 
condensation. 
 

While the equipment will not cause hazardous condition over this environmental range, 
performance may vary. 

Summary of Contents for SAS-550-1B

Page 1: ...A H Systems Model Active Monopole Antennas A H Systems inc REV E 1 Active Monopole Antennas User Manual...

Page 2: ...Monopole Antennas A H Systems inc REV E 2 TABLE of CONTENTS Introduction 3 Safety Instructions 4 Specifications 5 Operating instructions 6 Calibration 8 Accessories 10 Maintenance 11 Troubleshooting 1...

Page 3: ...r Description SAS 550 1B 9 kHz 60 MHz 2380 Active Monopole Antenna Battery Powered SAS 550 2B 100 Hz 60 MHz 2381 Active Monopole Antenna Battery Powered Included parts Antenna Base with ground plane T...

Page 4: ...cordance with all applicable laws and regulations 4 To protect against electric shock do not immerse the power cord power plug or Antenna in water or in any other liquid 5 Do not operate the product w...

Page 5: ...compatible with the receiver impedance ANTENNA SPECIFICATIONS SAS 550 1B Active Monopole Antenna specifications Frequency Range 9 kHz 60 MHz Antenna Factor 0 dB m Flatness 0 5 dB from 20 kHz 30 MHz 5...

Page 6: ...will not damage the battery Using the battery charger to operate the amplifier directly is not recommended Longest life and best performance can be obtained if the battery is charged when the air tem...

Page 7: ...des good reception with a minimal need for space Its flat 0 dB antenna factor makes it a convenient choice when calculating the field intensity The signal field strength is the receiver voltage in dB...

Page 8: ...the monopole element adjusted to 1 04m m a average radius of the antenna element 6 17mm m wavelength in meters Ca capacitance of the dummy load NOTE Other fixtures with different capacitance are avail...

Page 9: ...nal voltage VD in dB V at the T connector A Also referred to as a Dummy Antenna Figure 2 2 Leaving the RF output of the signal generator unchanged transfer the 50 termination to the T connector A and...

Page 10: ...able assembly providing repeatable reliable performance for use in a variety of commercial and industrial applications SAC 461F Cable kit for MIL STD 461 An optional cable kit is available to comply w...

Page 11: ...ll soft brush REPLACING THE BATTERY ASSEMBLY Turn off and unplug the unit before changing the battery pack assembly 1 Remove the four 6 32 flat head screws from the bottom of the ground plane 2 Discon...

Page 12: ...you still carry this part A Yes the telescoping rod with a BNC connector P N 2337 1 is for our older active monopole antennas All new active monopole antennas have been re designed to accept a telesco...

Page 13: ...tery pack will not charge Bad battery pack Replace battery pack assembly Charger has no output Re connect charger Charger not plugged in Replace battery pack assembly Low response No gain Front end da...

Page 14: ...the warranty period commencing with the delivery of the product by A H Systems The remedy set forth herein shall be the only remedy available to the buyer and in no event shall A H Systems be liable...