Installation location
The suction system should be located preferably in a clean, dry and well ventilated area.
Please be sure not to block the ventilation ports located on the motor housing. The unit
should be placed where the surrounding temperature remains between 10
o
C and 40
o
C
(50
o
F and 104
o
F). Always check to insure the location chosen is protected from direct or
indirect moisture contact.
Place suction outlet valve higher position than drainage to prevent backflow problem.
ATTENTION:
The maximum allowed distance between the suction and the dental chair is
6m for RAIN1 and RAIN1C and 8m for RAIN2.
Hose connection
Diagram of Suction tubing of RAIN1/RAIN1C
Diagram of Suction Tubing of RAIN2
Please be sure not to be bended any of suction inlet hose and outlet hose. This can
causes loosing vacuum power or damage of suction motor. Also use exhaust air hose with
40mm diameter hose.
Exhaust air temperature increases significantly and discharge drastically too hot for most
plastic piping. Therefore, metal piping or heat-resistant pipe is recommended for at least
the first five (5) to eight(8) feet (1.5 ~ 2.5 meters) from the air exhaust valve on the
discharge side. In addition, this piping MUST be guarded and marked
“DANGER–HOT–
DO NOT TOUCH.”