TURBINE Ti-940L
Please carefully read this Instruction for Use before use.
TURBINE Ti-940L is manufactured from the best materials, with a meticulous fi nish, a pleasant design and reliable performance. The products are
manufactured by combining German Technology with bearings from well-known brand bearing to create high performance.
TURBINE Ti-940L can be compatible with Kavo
®
MULTIfl ex
®
LUX, Sirona
®
LED, W&H
®
Roto Quick Coupling, Bien-Air
®
Unifi x
®
LED Coupling,, NSK
®
LED, Coupling, and will optimally meets the dentists’ requirements.
1. Proper use
The medical device is only intended for dental treatment: removal of carious, cavities and crown preparation, removal of fi llings materials, processing
of tooth, restoration of surface and surgical procedure through the rotary processing by the compressed air to drive the milling cutter at high speed.
CAUTION
– The medical device is intended for use by a qualifi ed knowledgeable user in dental workstation and laboratories.
– The medical device is intended for dental treatment. Any other type of use or alteration to the medical device is impermissible.
– Read carefully the Instructions for Use prior to using the medical device as it is always working at high speed during operation.
– During operation it is responsible for the user to protect the patient from danger and prevent contamination from the medical device.
– When beginning to run the medical device the user fi rst should test run away far from the patient to confi rm if it runs without any vibration,
overheat and noise, otherwise the use must stop running.
– When beginning to run the medical device the user should take protective measures.
– During operation the user must stop running if an abnormality in the medical device occur.
– A damaged cartridge will make disturbing noise to cause hearing loss of the patient; it is impermissible to use the medical device with the
en
damaged cartridge.
– It is impermissible to disassemble or remake the medical device without authorization.
– It is always to prevent the medical device falling down the ground or being hit by something hard.
– When not in use a test bar (delivered with the medical device) should be inserted in the chuck system.
– As the medical device is non-sterile it should be sterilized in autoclaves before being used.
2. Product description
Type KAVO
®
Multiflex
®
LUX
Type Bien Air
®
Unifix
®
LED
Type W&H
®
Roto Quick
Type Sirona
®
LED