Turning on
1. To turn on the Edge, press the
on/off
enter
key.
The start-up screen will appear
indicating the model and firmware
version. Then, the
measurement
view
will display.
Resetting
Resetting the dosimeter is performed
using QSP-II software via the clear tab
from the Edge Set-up button.
The Edge may also be reset using the
keypad. (See Chapter 3 and Chapter 4 in
the Edge user manual for details.)
Navigating
1
. Press the
up/down
arrow key to
navigate through measurement values,
battery condition, and run time.
Setting-up the Edge
All the Edge parameters are set up via
QSP-II software. (See
Chapter 3 in the Edge user manual for
details).
Calibrating
1. Remove the windscreen.
2. Ensure The Edge is turned on and
stopped.
3. Slide the Edge into the mouth of the
calibrator (ensuring the adapter ring is
inserted into the calibrator).
4. Turn on the calibrator. Ensure the
calibrator is set to 114 dB and 1000 Hz.
See subsequent steps on backside.
1
2
3
4
5
6
7
1. Windscreen & Microphone
2. LED indicator
Red
– indicates dose has
exceeded the alarm level.
Amber
– indicates
stopped.
Green
– indicates
measuring.
3. Measurement display
4. Up arrow
5. Run/Stop key
6. On/Off Enter key
7. Down arrow