Turning on 

1. To turn on the Edge, press the 

on/off

 

enter 

       key.   

 

 

 

The start-up screen will appear 

indicating the model and firmware 

version. Then, the 

measurement 

view

 will display.    

 

 

 

Resetting  

Resetting the dosimeter is performed 
using QSP-II software via the clear tab     

            from the Edge Set-up button.  

The Edge may also be reset using the 
keypad.  (See Chapter 3 and Chapter 4 in 

the Edge user manual for details.)

 

 

 

Navigating

 

1

.  Press the 

up/down 

arrow key to 

navigate through measurement values, 

battery condition, and run time. 

 
 

Setting-up the Edge

 

All the Edge parameters are set up via 
QSP-II software.                  (See 

Chapter 3 in the Edge user manual for 
details). 

 

Calibrating 

1. Remove the windscreen. 

 

2. Ensure The Edge is turned on and  

stopped. 

 

3. Slide the Edge into the mouth of the 

calibrator (ensuring the adapter ring is 
inserted into the calibrator). 

 

4. Turn on the calibrator.  Ensure the 

calibrator is set to 114 dB and 1000 Hz.  

 

 

See subsequent steps on backside. 

 

 

 

                 

 
 
 
 
 
 
 

 
 

1.  Windscreen & Microphone 

2.  LED indicator 

 

Red

 

– indicates dose has      

      exceeded the alarm level. 

 

Amber

 – indicates 

stopped. 

 

Green

 –  indicates         

     measuring. 

3.  Measurement display 

4.  Up arrow 

5.  Run/Stop key 

6.  On/Off Enter key 

7.  Down arrow 