7
English 202036A
Check the 3M
TM
Bair Paws
TM
system website to ensure you have the most recent version of this document.
www.bairpaws.com reorder #202036A
Preparing the Bair Paws
Model 875 Warming Unit for Use
Before using the warming unit, make sure that it is safely placed on a flat, hard surface such
as a table, or securely mounted on a wall, IV pole, suction wall mount adapter, or bedrail�
CAUTION:
Do not place the warming unit on a soft or uneven surface, such as a bed; this
may cause the air intake to become blocked, compromising the unit’s performance�
A wall-mount bracket is shipped with every warming unit; the brackets for mounting the
warming unit to an IV pole or a bedrail are available separately� Please contact your local
sales representative or call 3M Patient Warming Customer Service at 1-800-228-3957 for
more information about ordering mounting brackets�
Installing the Wall-mount Bracket
1� Using the four wall-screw holes on the wall-mount bracket (see Figure 1) as a template,
mark the position for the four wall anchors on the desired wall� The use of a level is
recommended�
2� Install four suitable wall anchors at the marked positions� If using the provided self-
drilling wall anchors, use a Phillips screwdriver to screw the anchors into position� If
installing into industrial-grade plasterboard, predrilling may be required�
3� Hold the wall-mount bracket in position against the wall�
4� Screw the four provided screws through the wall-mount holes and into the wall anchors�
Figure 1� Front view of wall-mount bracket�
Wall screws