3E RADAX SM 5000 Operation Manual Download Page 1

Features : 
RADAX  SM 5000 

 is a Digital Handheld Radiation Survey meter to detect  Gamma  and  X-

ray radiation.  
With 

RADAX SM 5000 

you can:

w

Detect  and  Monitor  Gamma  radiation  from  Radio-active  materials  and  X-ray 
machines. 

w

Check for leakage radiation

RADAX SM5000 

uses a Geiger-Mueller (GM) detector tube to detect radiation. The GM 

Tube contains ionization gas which generates pulse of electrical current when  radiations 
pass  through  the  gas  and  causes  ionization.  These  electrical  pulses  are  counted 
electronically and are termed as counts per minute (CPM) or they are displayed in milli-
roentgens per hour (mR/Hr). In SI units, the pulses are counted per second ( CPS) or micro-
Sieverts per hour (µSv/Hr).

RADAX  SM5000 

displays  radiation  level  detected  in  both  mR/hr  and  µSv/hr, 

simultaneously on its LCD display. The Digital survey meter is based on micro-controller 
technology and simple to use.

Technical Features : 

Detector

:

G M Counter

Radiations detected

:

Gamma rays & X-rays 

Energy response

:

100 Kev to 1.3 Mev

Range

:

0 - 5000 mR/Hr  & 0- 50000 µSv/Hr (Auto-Ranging)

Accuracy with Cs137

:

± 15 %

Response time

:

15 Sec to 20 Sec.

Operating Conditions

:

0- 50®C 90% Humidity

Display

:

Digital - LCD - 2 lines.

Battery

:

2 nos. AA size - Alkaline.

Controls

:

Touch Switch for  Power On-Off / Function

Construction

:

ABS Plastic

Operation :

There are two touch switches on the Survey meter. Power & Enter switch and Setting switch. Press Power switch and keep 
pressed till display is on. Display will show reading in mR/Hr & µSv/Hr. Bring the Survey-meter near the Radiation source 
with the Detector sign pointing towards the Radiation source. The survey meter will display the reading in both mR/Hr and 
µSv/Hr.   
While detecting the radiation, the audio buzzer gives out chirping sound. At low level radiation, the frequency of chips is 
low. As the radiation level increases, the chirps become more frequent. At very high radiation or very near the source, the 
chirping turns in to a whistling sound. 
The buzzer can be switched off when required. Press the Setting switch and keep pressed. Select option Buzzer On, Press 
Power/Enter switch to save the setting. To switch off Buzzer, select Buzzer Off option. To switch off the Survey meter, press 
power switch and keep pressed till Power Off sign appears.

RADAX

DIGITAL RADIATION SURVEY-METER

OPERATION MANUAL

MODEL :  RADAX SM 5000 

300 Hwy 146 N, La Porte, Tx 77571

Ph # (281) 470-2010, Fax # (281) 470-2024, Cell # (954) 895-1603 

Reviews: