PANNORAMIC® CONFOCAL RX 1.0
OPERATING SLIDES AND TRAYS
MAY 13, 2021
28
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
The label area is recorded by the preview camera and can be displayed by the Pannoramic®
Scanner Software for research control software and viewer software.
To affix barcode stickers to slides, do the following:
1)
Peel off the barcode sticker
(2)
from the roll.
2)
Stick it on the colored label area
(1)
of the slide, on the specimen side.
Caution!
Ensure that a marginal space of 1-2 mm remains on all sides between the sticker and the
label area outer limits. It is vital for barcode identification. Do not let barcode sticker
protrude from slide edges or stick on the coverslip surface
(3)
.
The thickness of the barcode sticker on the slide must not exceed 0.4 mm!
Do not stick more than
four
barcode stickers over each other and the overall thickness must
must be kept below 1.65 mm (including the barcodes and the slide).
3.3
Inserting and removing slides
To insert slides into a tray
Figure 7: Inserting slides into a tray
A tray can contain a maximum of 11 histological slides. In the 12
th
position a
Cleaner slide
must be
loaded.
•
Push a slide (
1
) into the desired slot (
4
) of the tray (
3
) with its barcode end (
2
) first
.
•
Make sure that the slides are inserted correctly in the tray.