PANNORAMIC® DESK FLASH DX
OPERATIONAL SAFETY
20
INFORMATION FOR USE
–
FOR INVESTIGATIONAL USE ONLY.
THE PERFORMANCE CHARACTERISTICS OF THIS PRODUCT HAVE NOT BEEN ESTABLISHED.
Liquids and Safety
Warning!
Prior to cleaning the Scanner, the power cord should be unplugged. Only use a dry or lightly
damped lint-free cotton cloth to clean the Scanner. Isopropanol is recommended as a
cleaning agent to clean greasy or dirty surfaces. Make sure that the affected area is
completely dry before switching the device on and operating the system again.
Security
Access control
User name and Password on the user interface must be entered to access and enable full
functionality of the control software.
Caution!
The
Pannoramic® DESK FLASH DX
cannot be operated without user authentication during
login to the control software.
Caution!
The automatic logout function of the
Pannoramic®
Diagnostic Scanner Software prevents
the device from unwanted access by logging the user out automatically after the specified
time of inactivity.
Patient and data security
Warning!
Take extra care when labeling or inserting slides into the magazines as such mistakes can
result in skipped slides or unidentified slides in CaseManager, or lead to patient data mix-up.
For proper labeling of glass slides, the instructions in chapter
Operation
must be followed.