DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp.  Page 1 of 4 

GCX Mounting Assembly Operation/Installation Manual 

Dräger Model Cato

 ®

 with Siemens SC9000 Patient Monitor 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

Patient Monitor – Left Side on Pivot Arm  /  Power Supply on Right Side 

 

The purpose of these instructions is to: 

 

 

1.  Describe attachment of GCX mounting hardware to the Cato anesthesia machine. 

 

2.  Describe attachment of the Siemens monitor to the GCX mounting hardware. 

 

WARNING

:  USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE 

    DESCRIBED 

IN 

THIS 

DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE  

    ANESTHESIA 

MACHINE. 

 

 

 

 

 

THIS CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH  

    EXTENDED 

WHEEL 

ADAPTER 

UPGRADE 

SET FROM DRÄGER (PART NUMBER:  86 018 68) 

 

Installer

:  When installation is completed, provide these instructions to the end user for future reference. 

 

INSTALLATION 

 

Tools Required:  Philips head screw driver, Hex key (incl.). 

 

Mounting the Channel to Cato Edition 

 

Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2). 
     Brass 

Plate 

with 

inserts 

from 

Dräger 

(M29110) 

 

1.  To attach the Channel to the Cato left side accessory track,  
 

locate the brass plate in the accessory track, align the  

 

mounting holes in the Channel with the threaded holes on  

 

the brass plate then run the M5 flat head screws (2) partly  

 

in. (Note:  Fixed Stop is at the bottom of the Channel) 

2.  Set the height of the Channel in the accessory track and  
 

tighten the M6 screws to hold in place. 

 
Prepare M Series Arm 

 

1)  Install the Tilt Limiting Stop screw. 
 
 
 
 
 
 

M6 x 16 mm (2) FHMS 

Accessory Track

Fixed Stop 

5/16-18 x ½” SHCS (1)

 

Summary of Contents for SC9000

Page 1: ...D WHEEL ADAPTER UPGRADE SET FROM DRÄGER PART NUMBER 86 018 68 Installer When installation is completed provide these instructions to the end user for future reference INSTALLATION Tools Required Philips head screw driver Hex key incl Mounting the Channel to Cato Edition Parts Required 4 Channel M6 x 16 mm FHMS screws 2 Brass Plate with inserts from Dräger M29110 1 To attach the Channel to the Cato...

Page 2: ...m of the Pick Go base with the edge of Mounting Plate Slide the Pick Go base into the plate until it stops at the Spring Plunger located below the Mounting Plate Pull the Spring Plunger and continue to slide the base until the mounting adapter engages the Spring Plunger Tighten the 4 nylon thumb screws on the back of the Mounting Plate Install 4 Cato Channel on Right Side As described on Page 1 of...

Page 3: ...evice 2 To adjust the swivel tension use a or 13mm socket driver to tighten or loosen the Swivel Tension Adjustment Bolt located under the front end of the Arm Tilt 1 Grasp the top and bottom of the device and tilt it to the desired angle 2 To adjust the overall tilt tension evenly tighten or loosen the 2 10 32 x 5 8 SHCS Tilt Tension Adjustment Screws using the supplied 5 32 4 mm hex key Cable Ma...

Page 4: ...pe any cleaning agents off of the arm immediately using a water dampened cloth Dry the arm thoroughly after cleaning CAUTION GCX makes no claims regarding the efficacy of the listed chemicals or processes as a means for controlling infection Consult your hospital s infection control officer or epidemiologist To clean or sterilize mounted devices or accessory equipment refer to the specific instruc...

Reviews: