SP

Service Instructions

From Serial No.

MOBILETT Plus

3200

MOBILETT Plus E

10900

66 08 595

Summary of Contents for 3200 Series

Page 1: ...ion or use of this document or its contents is not permitted without express written authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model _or_ design _are_ reserved SP Service Instructions From Serial No MOBILETT Plus 3200 MOBILETT Plus E 10900 66 08 595 ...

Page 2: ...els Disclaimer The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho rized by Siemens or one of its affilia...

Page 3: ...________________________________ 4 1 5 _______Service programs ______________________________________________ 5 1 Entering service mode 5 1 Selecting test programs with the kV buttons 5 1 Error messages in service mode 5 1 List of test programs available in service mode 5 2 P01 Filament test 5 2 P02 Rotation test 5 2 P03 Capacitor bank and charging test 5 3 P04 Display led and buzzer test 5 4 P06 ...

Page 4: ...t intensity 9 7 Light field relative to the X ray field 9 7 10 _____ Mechanical systems ___________________________________________10 1 Adjusting the arm system 10 1 Parallelism 10 1 Counterbalance 10 3 Adjusting the friction linings 10 4 Replacing the spring pack 10 7 Adjustment and replacement of the turn plate 10 11 Replacement of the single tank 10 14 Replacement of the support arm cable harne...

Page 5: ...ns marked with this symbol have been included in the Siemens Training Video part of the self study packet which can be ordered through Siemens customer service DANGER indicates when there is an immediate danger that leads to death or serious physical injury WARNING indicates a risk of danger that may lead to death or to serious physical injury CAUTION used with the safety alert symbol indicates a ...

Page 6: ...e oscilloscope with 2 5 accuracy Torque wrench 1 Nm 3 Nm Spring scale Lux meter Rope Pulley block Adjustment device for belt tension part No 65 64 301 Lubricants 1 All purpose grease PD 2 Viscogen oil Wiring Diagram RXR8 215 051 Function description SPR8 215 850 Installation Instruction Installation of remote exposure switch system SPR8 220 031 02 All tools measurement and auxiliary devices with t...

Page 7: ...107 or country specific regulations Remove or install boards only when the generator is switched OFF Adhere to the ESD guidelines Checks and adjustments performed with radiation ON are identified by the radiation warning symbol During these types of adjustments radiation protective clothing must be worn Never work with the system open if the capacitor is charging or has just charged The capacitor ...

Page 8: ...hen remove the left and right covers and the cover with the cassette compartment Fig 1 Fig 2 Comply with the information given in this section before opening the system Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may still be charged even if the system is switched OFF and the line voltage cable is disconnected Life threatening electric s...

Page 9: ...e check the measurement device Measure the voltage for the capacitor bank at the bus bar Fig 3 Only use insulated tools when pulling out the capacitor units If the voltage measured between CAPPOS and CAPNEG is 0V the measurement device could be defective or the wires between the CAPPOS and CAPNEG test points on the capacitor bank could be damaged The capacitors could still be charged In this case ...

Page 10: ...re the capacitor bank voltage on the bus bar Fig 3 If one of the capacitor series of the capacitor bank is still charged U 10 V it can be discharged via resistor R103 on D7 Connect the voltmeter to the measurement points parallel to the insulated cables which will be used to discharge the capacitor bank U 10 V Fig 4 ...

Page 11: ...eck the filament heating value for the selected kV see section P07 Read out filament values from battery backed RAM Compare the value with the value in the delivery protocol If the difference is large make a new adaptation see section P10 Fast adaptation Err 3 in combination with Err 21 Err 22 Check the measuring point FIL_VAL on D1 see page 160 and page 110 CAL 13 Check the 60 V supply voltage on...

Page 12: ... 141 For a large deviation Check the filament value see section P07 Read out filament values from battery backed RAM and compare it with the values in the delivery protocol If the difference is large a new adaptation is needed see section P10 Fast adaptation See also section Error diagnosis HT and single tank Check the connection K9 between D1 and H1 single tank and the connection between D1 and D...

Page 13: ...ce on the generator a defective connection may be the cause Check the connections K64 1 and K64 2 on D13 and the K64 connection to the single tank see page 140 Check cables K9 and K8 see page 140 Check main fuse U101 see page 140 Check the trigger pulses INV A and INV B to the inverter page 140 Check the capacitance in the capacitor bank see section P03 Capacitor bank and charging test Check the c...

Page 14: ... 140 during the next test Activate the radiation release switch to the first level preparation and listen for the anode to start up refer to page 150 Check that the following supply voltages are present on D4 100 V 15 V 5 V DC refer to page 150 Check that the trigger pulses are present at ROT when the anode is accelerated refer to page 150 Check that the measuring signals ROT_VAL on D4 and ROT on ...

Page 15: ...tank see section Error diagnosis HT and single tank Err 9 During exposure When measuring the kV value a large high tension asym metry is detected Normally the kV value is symmetric around ground The acceptable devi ation is 19 5 kV This corresponds to HTD 2 5 V 0 5 V Intermittent measurement cables or defective single tank Check the measurement cable K9 between D1 and the single tank refer to page...

Page 16: ...surement circuit for the temperature refer to page 140 contact K9 CAL 12 During preparation Stored filament value is lost Check the voltage in the backup battery refer to page 110 The battery should have a voltage 2 8 V Check that the battery jumper is securely attached see page 110 Make a new adaptation see section P10 Fast adaptation CAL 13 Generator ON The voltage supply 60 V DC to the fila men...

Page 17: ... D7 If the transistor in the Buck converter is short circuited the current will flow until fuse U8 on D8 blows Replace the D7 charging unit USE 15 During preparation The exposure switch has been in prepara tion position longer than 20 s The user has kept the exposure release switch in preparation position longer than 20 s after the Ready light indicator indicates that preparation is ready anode sp...

Page 18: ... the internal residual voltage in the modules was too high when they were connected to each other This caused the fuse to blow refer to page 130 Err 17 Generator ON The capacitors are charging too slowly 1 4 V s Check that S10 on D7 is switched off If the error appears after working on the capacitor bank check the cabling and the connections to the capacitor bank refer to page 130 Check fuse U8 on...

Page 19: ... glows The leakage current in the capacitor bank is too large The measurement is performed dur ing charging at 150 V DC Charging is interrupted for 5 s If the voltage drops more than 1 6 V a message will be displayed See also Err 17 In cases where the generator has not been operated for several months the leakage current in the capacitors has increased Use the formation program P06 refer to sectio...

Page 20: ...c tions k11 and k1 refer to page 140 Err 21 Generator ON The measured value of the stand by fila ment current has exceeded the limit FIL_VAL on D1 shall be within 0 16 2 5 V when the filament is in stand by mode Check the supply voltages on D4 60 V 15 V and 5 V refer to page 160 Check the pulses at measurement point FIL on D4 refer to page 160 Check the value in FIL_VAL on D1 refer to page 160 Che...

Page 21: ... refer to page 160 Check the connection between D4 and the filament transformer K60 refer to page 160 Check the trigger pulses in measurement point FIL on D4 refer to page 160 Check the value in FIL_VAL on D1 The value must be stabile in stand by filament After change over to preparation the value should stabilize again Refer also to the section P01 Filament test Dis played messages Mode of operat...

Page 22: ...erter current refer to the section Error diagnosis HT and single tank The X ray tube may have arced during the exposure If the first exposure after a long interval is an exposure with high kV values tube arcs can occur The unit has to be started with several exposures at high mAs values The kV values selected should be from 40 kV increased to 133 kV Refer also to the section Error diagnosis HT and...

Page 23: ...p battery with battery jumper are okay replace D1 Refer to page 110 and 140 USE 25 Generator ON The calculated load of the anode target has been lost The value is lost from CMOS mem ory See also CAL 31 Check all supply voltages on D1 see page 110 Check that the battery jumper is properly mounted and that the vol tage in the backup battery is 2 8 V refer to page 110 A tube arcing has caused a tempo...

Page 24: ... CAL 26 occurs repeatedly or if the message can not be acknowledged by pressing kV or switching the unit OFF and ON replace D1 CAL 27 Generator ON during start up Error in ROM mem ory Check sum is not correct Check that all supply voltages on D1 are present refer to page 110 The ROM containing the program is defective Replace the software IC s I72 I66 and I74 on D1 refer to page 110 If the error o...

Page 25: ...ld be replaced Enter the cause and the error message CAL 28 in the service protocol CAL 29 Generator ON The single tank is not correctly con nected The single tank is missing or the cable harness is incorrectly connected Check that connection K9 on D1 is connected Check for possible damage in the main cable harness between D1 and the single tank Measure K9 13 4 and move the arm and single tank whi...

Page 26: ... has disturbed the system If the USE 30 message occurs repeatedly the single tank must be checked according to Error diagnosis HT and single tank If the function of the single tank is OK and USE 30 appears repeatedly replace D1 If necessary reinstall the customer kV and mAs limit values Refer to the section P14 Number of kV steps and upper kV limit and P15 Upper mAs limit CAL 31 Generator ON The v...

Page 27: ...signal fault during the pro gram execution Err 33 messages can be ignored This message can only be read out from the message history P08 The message is only used as aid for the software development Acknowledge and clear the mes sage history Err 34 The mes sage is not shown to the opera tor This message can only be read out from the mes sage his tory P08 Generator ON The number of mes sages is beyo...

Page 28: ...e can only be read out from the mes sage his tory P08 Generator ON The inverter was interrupted once e g overcurrent If the inverter has been stopped more than 3 times the exposure will be interrupted and the Err 23 mes sage is shown on the display Check the inverter current refer to the section Error diagnosis HT and single tank Refer also to P13 Inverter test The X ray tube may have arced during...

Page 29: ...ly the single tank must be checked according to section Error diagnosis HT and single tank If the function of the single tank is OK and USE 37 appears repeatedly replace D1 Err 38 The mes sage is not shown to the opera tor This message can only be read out from the mes sage his tory P08 Generator ON The error statistics are lost See Err 37 If Err 38 is the only message in the history memory it wil...

Page 30: ...kVactual value results in decreased heating value negative correction 2 Change in generator output This may occur when exchanging the inverter or the single tank In each case the heating values must be readjusted and recorded refer to P10 and P07 in the service program a b If the generator output is increased this results in an increase in heating value If the generator output decreases the heatin...

Page 31: ...harging too fast Err 17 Capacitor bank charging too slow Err 18 Capacitor voltage charging too slowly CAL 19 Leakage current too high in the capacitor USE 20 Radiation release button was released during the exposure Err 21 Filament current incorrect for stand by Err 22 Filament current incorrect for exposure Err 23 During exposure the main inverter was interrupted 3 times USE 24 The selected kV an...

Page 32: ...Siemens Elema AB Rev 01 01 02 SPS UD Solna Sweden 2 22 Display messages List of messages in the error memory 2 Err 33 Signal fault during program execution Err 36 Overcurrent in HT inverter Err 37 Error memory contents lost Err 38 Error statistics lost ...

Page 33: ...under A E in the sec tion Troubleshooting instructions Power plug inserted MOBILETT Plus MOBILETT Plus E ON Display illuminated kV and mAs buttons OK Collimator lamp can be turned ON OFF Turn plate can be activated deactivated Green Ready light indicator is ON Voltage supply Program run see section P04 Display led and buzzer test Connections Button function Defective lamp Cables Fuse Lamp button r...

Page 34: ...voltage Exposure Exposure interrupted Correct density MOBILETT Plus MOBILETT Plus E OK Yes Yes Yes Yes No Yes Exposure release switch cables and connections Err 6 see P02 Rotation test Remote control option Exposure release switch cables and connections Displayed message Remote control option kV mAs selection Exposure time Heating value Darkroom No USE 39 Err 6 Err 6 No No No Displayed message No ...

Page 35: ...he voltage supply on D4 Check the measuring point LAMPON If there is no signal check k5 k9 k3 and k2 If a click is heard check the voltage supply in the lamp bracket refer to the wiring diagram page 180 MOBILETT Plus only D Press the turn plate button and check if a click is heard from MR1 and MR2 refer to the wiring diagram page 170 If a click is heard check the voltage in K68 1 2 and compare it ...

Page 36: ...MOBILETT Plus Plus E SPR8 215 898 02 Page 2 of 2 Siemens Elema AB Rev 01 01 02 SPS UD Solna Sweden 4 2 Troubleshooting instructions This page intentionally left blank ...

Page 37: ...s displayed pressing the kV button will show the program version again and so on Pressing the kV button will display the previous test program The lamp button has a toggling function and is used both to start and stop the displayed test program For exam ple after the lamp button has been pressed the formation of the capacitor bank starts and the following is shown on the display The test program c...

Page 38: ...m X037E page 150 Accelerates the anode with a 160 Hz control signal during two seconds After this first acceleration the speed is measured and if necessary a second acceleration of two sec onds is made When the correct speed is reached the anode will rotate freely for 20 sec onds or until the rotation frequency is below 142 Hz before it is braked during 9 seconds to a stop An error message will be...

Page 39: ...pacitor will first be charged to 350 V then discharged to 229 V The voltage will be held for 5 seconds before it is restored to 350 V During the test the voltage in the capacitor bank is indicated except when the capacitance is displayed If an error is detected the test program will be aborted An audible beep will be signaled and a message will be displayed If the test ends successfully the displa...

Page 40: ...ank Approximately every six minutes the voltage value is increases The charging sequence increases as follows 150 250 299 325 337 345 348 351 353 V At the 353 V level the charging will continue for another 48 minutes to decrease and stabilize the capacitors internal leakage current If an error is detected during the formation program the program is aborted When the program is aborted there is an a...

Page 41: ...re sponding message will be displayed in the mAs field Pressing the kV button will dis play the next higher number and corresponding message and vice verse with kV The program starts by showing the latest message in the history If there are no messages the display will show The latest message has the highest number for example To exit P08 use the lamp button P09 Clear the message history 5 Activat...

Page 42: ...ed since the generator has incorrect filament values However check every message against the message list and confirm them by pressing the kV button When the generator is ready with an adaptation on a kV value it continues to the next value When the adapta tion is finished the program will stop and the display will show P11 Backup battery test 5 Wiring diagram X037E page 110 After activating the p...

Page 43: ... values and activates the second part The second part of P14 is the selection of the upper kV limit The display shows the selected highest available kV min 40 and max 133 kV This can be changed with the kV or buttons Example Pressing the lamp button stores the currently displayed values and also exits the program P14 P15 Upper mAs limit 5 The program enables you to set the upper mAs limit The avai...

Page 44: ... lamp button and shows the actual number of exposures Example 1 Example 2 Exit the program by pressing the lamp button kV mAs 599 The number of exposures is 599 kV mAs 1 15055 The number of exposures is 115 055 The exposure counter can not be cleared via any of the service programs The value is stored in the CMOS memory If the backup battery is depleted the value could be lost NOTICE ...

Page 45: ... cannot be displayed check the tube current at measurement point MAPOS on D1 Check the inverter current with an external current transformer Measure HTPOS and HTNEG on D1 Record the fall times 90 10 of HT high voltage levels and whether the exposure was interrupted prematurely Compare with fig 2 HT signal inverter and single tank normal Measure the inverter current using an external current transf...

Page 46: ...is less than 100 ohms the inverter is defective transistors in the modules are bad The inverter current is always measured using a separate current transformer between the inverter and single tank The procedure is described in the training video Measuring the falling edge of the high voltage in the measurement point HT on D1 board Long falling edge approx 2 ms Measuring the inverter current Separa...

Page 47: ...t normal High speed HT and HT discharges Figure 6 Defective tube Defective high voltage capacitor Arcing between plus and minus side Replace single tank Abnormal inverter current e g triangle shaped HT dis charges Figures 9 and 10 Defective high voltage transformer Defective diode strip Defective tube Replace single tank Abnormal inverter current e g asymmetric Figure 10 Defective inverter Replace...

Page 48: ...a AB Rev 01 01 02 SPS UD Solna Sweden 6 4 HT and single tank error diagnosis Exposure values 90 kV 2 0 mAs HTPEAK 20 kV div Inverter current 100 A div Fig 1 HTPEAK and inverter current inverter and single tank normal HTPEAK on D1 Inverter current Inverter current ...

Page 49: ...age 5 of 12 MOBILETT Plus Plus E Solna Sweden Rev 01 01 02 SPS UD Exposure values 40 kV 5 mAs HT 8 kV div 200 mV div Iinverter single tank 100 A div only the end of the exposure is shown Fig 2 HT and inverter current inverter and single tank normal Inverter current HT on D1 ...

Page 50: ... 01 02 SPS UD Solna Sweden 6 6 HT and single tank error diagnosis Exposure values 90 kV 5 mAs HT 20 kV div 500 mV div Iinverter single tank 50 A div only the end of the exposure is shown Fig 3 HT and inverter current inverter and single tank normal HT on D1 Inverter current ...

Page 51: ...Page 7 of 12 MOBILETT Plus Plus E Solna Sweden Rev 01 01 02 SPS UD Exposure values 33 kV 5 mAs HT 40 kV div 1 V div Iinverter single tank 100 A div only the end of the exposure is shown Fig 4 HT and inverter current inverter and single tank normal HT on D1 Inverter current ...

Page 52: ...Elema AB Rev 01 01 02 SPS UD Solna Sweden 6 8 HT and single tank error diagnosis Exposure values 133 kV 5 mAs HTPOS HTNEG 2 V div 20 kV div 200 µs div only the end of the exposure is shown Fig 5 HT signal inverter and single tank normal HTPOS on D1 HTNEG on D1 ...

Page 53: ...9 of 12 MOBILETT Plus Plus E Solna Sweden Rev 01 01 02 SPS UD Exposure values 133 kV 5 mAs HTPOS HTNEG 2 V div 20 kV div 200 µs div only the end of the exposure is shown Fig 6 HT signal Arcing between plus and minus side Defective tube Replace single tank HTPOS on D1 HTNEG on D1 ...

Page 54: ...D Solna Sweden 6 10 HT and single tank error diagnosis Exposure values 133 kV 5 mAs HTPOS HTNEG 2 V div 20 kV div 100 µs div only the end of the exposure is shown Fig 7 HT signal high speed HTNEG discharges Defective tube filament transformer wires to the cathode HTPOS on D1 HTNEG on D1 ...

Page 55: ...en Rev 01 01 02 SPS UD CH1 20 mV div 100 A div 10 µs div Inverter current CH2 500 mV div 20 kV div HT CH1 Inverter current 50 A div 10 µs div Fig 8 High speed HT discharges Defective tube transformer capacitor Fig 9 Abnormal inverter current Defective single tank CH1 CH2 Inverter current HTPOS on D1 T Inverter current ...

Page 56: ...en 6 12 HT and single tank error diagnosis CH1 Inverter current 50 A div 20 µs div CH1 I V div 40 kV div 20 µs div HT CH2 100 A div Inverter current Fig 10 Abnormal inverter current Defective inverter Fig 11 Abnormal inverter current Defective inverter Inverter current T HT on D1 Inverter current T ...

Page 57: ...or malfunctions the internal series fuse will blow and the capacitor series will be disconnected Depending on the tolerance in the capacitors the generator will display a message If a large positive tolerance is present in the capacitors the balance of the capacitor bank will be able to maintain the total capacitance above the lower limit Both the capacitor and the fuse must be replaced according ...

Page 58: ... 10 V Fig 2 If the voltage is measured to 0 V the measuring device could be defective or the cables between the capacitor bank and CAPPOS CAPNEG could be damaged In this case proceed as described in items a e below to avoid personal injury Fig 1 Fig 2 a Connect the power plug b Connect the measurement device c Switch on S10 on D7 Check whether the voltage increases d Switch off S10 Check whether t...

Page 59: ...en Rev 01 01 02 SPS UD Measure the voltage between the and bars the voltage must be less than 10 V Note the cable connections to the capacitor bank Fig 3 for MOBILETT Plus and Fig 4 for MOBILETT Plus E Fig 3 MOBILETT Plus Fig 4 MOBILETT Plus E ll MOB00613 bar neg bar pos MOB00614 bar neg bar pos ...

Page 60: ...rs Fig 5 for MOBILETT Plus and Fig 6 for MOBILETT Plus E Remove the capacitor module Fig 5 MOBILETT Plus Fig 6 MOBILETT Plus E Fig 7 Residual voltage up to 353 VDC may still be present Risk of high voltage Use insulated tools and do not touch bars that may contain volt age Fig 7 MOB00496 bar pos bar neg MOB00495 bar pos bar neg WARNING ...

Page 61: ...ane the capacitor must be replaced If you cannot determine the error visually perform tests A D on the capacitor series with the blown fuse Fig 8 Electric shock hazard Risk for physical injury Prior to installing a new capacitor or a new fuse ensure that the voltage difference between capacitor series or capacitor modules does not exceed 2 volts A Remove the fuse and bar in the defective series Wa...

Page 62: ...ll screws and nuts on the upper side of the capacitor module with a torque of 3 Nm and all screws and nuts on the under side with a torque of 1 Nm Reattach the left and right covers and the column cover Connect the power plug and switch off S10 on D7 Switch on S 1 1 on D1 Read the information about the formation program P06 in section P06 Capacitor formation if you have replaced one or more capaci...

Page 63: ...nter Read out the mAs limit refer to the section P15 upper mAs limit Read out the kV step and limit refer to the section P14 Number of kV steps and upper kV limit Loosen the screws to the system upper cover and remove it Loosen the screws and remove the protection plate Locate the battery refer to the wiring diagram page 110 To avoid damage to components or assemblies ensure compliance with ESD gu...

Page 64: ...g ON S 1 1 on D1 Perform the adaptation according to section P10 Fast adaptation Set the selected mAs limit refer to the section P15 Upper mAs limit Set the selected kV step and limit refer to the section P14 Number of kV steps and upper kV limit Read out the exposure counter refer to the section P16 Show the exposure counter Switch off S 1 1 on D1 OFF Reattach the system upper cover Make an expos...

Page 65: ...the cassette compartment Remove the four two on each side screws that fixes the cable winder assembly pos 1 Fig 2 Pull out the cable winder assembly approximately 10 cm Disconnect main switch plug K52 pos 1 Fig 3 Fig 2 Fig 3 Sharp edge The brake handle could be damaged Lock the brake handle otherwise it is damaged by the sharp upper edge of the cable winder assembly MOB00618 1 MOB00619 1 CAUTION ...

Page 66: ...able ties pos 1 Fig 4 Remove four screws and four nuts pos 2 Fig 4 and Y plate pos 3 Fig 4 Disconnect cable from upper plate pos 1 Fig 5 Remove the distance ring pos 2 Fig 5 from the axle Loose the brake carefully to loose the tension of the cable winder spring Remove securing screw pos 1 Fig 7 Pull out the cable winder Fig 4 Fig 5 MOB00620 2 1 3 4 MOB00621 1 2 3 3 ...

Page 67: ...nder has to be replaced go to B If mains cable has to be replaced Remove four screws pos 3 Fig 5 Remove connection plate and disconnect the mains cable pos 1 Fig 6 Install new mains cable and connection plate B Reinstall ring cable stop and plug Reinstall cable winder Reinstall the securing screw pos 1 Fig 7 Fig 6 Fig 7 Wind the mains cable to the cable winder without any play 1 MOB00622 1 NOTE ...

Page 68: ...winder spring Repeat that six times Reinstall upper plate with distance ring plastic role and Y plate Pull out the cable completely Test cable winder and brake function Reinstallation is performed in the reverse order Fig 8 The cable winder should not rewind by itself The brake force is adjusted via right screw nut clockwise stronger pos 1 Fig 8 nut counterclockwise softer MOB00620 1 NOTE ...

Page 69: ... collimator lamp by loosening not removing the screws marked B one turn Do not adjust screws C D or E Fig 2 Remove the defective lamp and replace it with a new lamp Fig 3 Do not touch the bulb with bare fingers See list of spare parts Reinstall the protective plate Tighten the screws marked B Fig 1 Fig 2 Fig 3 Equipment required Lux meter 75 75 75 75 90 90 75 75 75 75 60 60 45 45 30 30 15 15 0 15 ...

Page 70: ... measurements made in the approximate center of each quadrant of the light field Fig 4 If the illumination is less than 180 lux check the lamp supply Refer to wiring diagram X037E page 180 Light field relative to the X ray field 9 Load a 35 x 35 cm 14 x 14 cassette with film Put the cassette on a table or a similar surface Fig 4 Fig 5 Fig 6 Equipment required Lux meter Equipment required Cassette ...

Page 71: ...0 mAs on the control panel Release an exposure Develop the film and measure the deviation between the light and the radiation fields Acceptable deviation 1 7 of SID Fig 6 If the deviation is more than 1 7 adjust with the screws marked C and D Fig 2 to move the light field closer to the radiation field Repeat the test with a new exposure If necessary adjust again until the deviation between the lig...

Page 72: ...nt and repairs on the collimator Replacing the collimator 9 Switch off the unit and disconnect the power plug Turn the single tank so that the collimator is facing upwards Fig 7 Secure the arm and the single tank with a rope Fig 7 Remove covers A and B Remove collimator cover C Fig 7 Equipment required Rope C A B ...

Page 73: ...ction terminal board D Fig 8 If there are cable ties cut them Fig 8 Remove the collimator by loosening the four screws E Fig 8 Reserve the hardware Install the new collimator Center the rotary flange so that the screws marked E are centered in the grooves before tightening Connect the wires to connection terminal board D Fig 8 D E E E E cable tie right side cable tie left side ...

Page 74: ...t for the left side but with the cable tie mirror inverted Rotate the collimator through the whole interval and make sure the cable ties does not come in contact with the collimator moving parts Reattach collimator cover C Make sure not to forget to attach the cable ties When replacing the collimator the new collimator should be positioned and adjusted so that the light and radiation fields corres...

Page 75: ...surements made in the approximate center of each quadrant of the light field Fig 11 If the illumination is less than 180 lux check the lamp supply Refer to wiring diagram X037E page 180 Light field relative to the X ray field 9 Load a 35 x 35 cm 14 x 14 cassette with film Put the cassette on a table or a similar repository Fig 11 Fig 12 Fig 13 Equipment required Lux meter Equipment required Casset...

Page 76: ...ins Fig 12 Set 52 kV 2 0 mAs on the control panel Release an exposure Develop the film and measure the deviation Acceptable deviation 1 7 Fig 13 If the deviation is greater loosen the Allen screws marked E slightly and move the rotary flange to desired position Tighten the screws marked E Fig 14 Repeat the test and adjustment of the rotary flange position until the deviation between the light and ...

Page 77: ...each other as possible but with the same diameter of the pull rod Fig 26 The parallelism must not deviate more than 0 6 mm The measurements must be performed with the arm in the uppermost and lowermost positions If not parallel loosen the locking nuts and adjust by turning the pull rod Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may stil...

Page 78: ...han 35 mm Fig 3 If the visible part of the adjustment rod is longer than specified the single tank and the arm might fall If any adjustment is needed the single tank should be secured in its uppermost position with a ceiling support Fig 2 After the single tank has been secured the two nuts can be loosened and the adjustment rod can be turned Lock the rod with the nuts and repeat the parallelism ch...

Page 79: ... in the column Fig 4 and the lining in the lower joint Fig 5 Check that the arm system stops in any position If does not stop but moves lower or higher Adjust the arm system with the adjustment screw Fig 6 until it balances in all positions If the arm moves lower the screw must be tightened if it moves up the screw must be loosened Fig 4 For upper arm stability Fig 5 For lower arm stability Fig 6 ...

Page 80: ...oles in the cover Fig 4 Loosen the screws completely with a 4 mm Allen wrench Retighten the screws until there is no play Tighten the screws evenly until the upper arm holds its position but is still easy to move Lower arm If the lower arm does not hold its position it must be adjusted Fig 5 Loosen the screw completely Retighten the screw until there are no play Tighten the screw until the lower a...

Page 81: ...n it must be adjusted Fig 7 Loosen the screw completely Retighten the screw until there is no play Tighten the screw until the fork holds its position Movement of the single tank If the single tank rotates too easily or does not hold its position in the fork Remove the caps on the fork Tighten or loosen the friction adjustment one on each side of the fork Fig 8 Fig 7 Fig 8 MOB00580 ...

Page 82: ...2 Page 6 of 28 Siemens Elema AB Rev 01 01 02 SPS UD Solna Sweden 10 6 Mechanical systems Movement of the collimator Remove the collimator cover Adjust the collimator movement resistance with the screw 1 Fig 9 Reattach all covers Fig 9 1 ...

Page 83: ...r Remove the upper column cover by removing the transport handle and the lower column cover by lifting it upwards Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may still be charged even if the system is switched OFF and the mains cable is disconnected Life threatening electric shock hazard exists See Chap Protective measures when working o...

Page 84: ... Disconnect all cables ground connections and K68 according to Fig 11 Remove the locking rings Remove the bearing rings Remove the shaft Let the turn plate down and remove it The ceiling suspension must be able to hold a weight of 100 kg since the entire weight of the arm will be on the ceiling support once the spring is removed If not the arm and single tank will fall Make sure the ceiling suppor...

Page 85: ...d and check through the hole for the adjustment screw that the front of the spring pack with recess is visible Mount the adjustment screw and tighten until the spring pack is held in position Fig 12 Risk of injury to the hands The strong tension in the spring pack could cause undesirable actions Be careful during the following procedures If the spring pack does not lower there is still spring tens...

Page 86: ...ngs Connect the ground cable and K68 All MOBILETT s Reinstall the supporting wheels Remove the ceiling support carefully and check that the single tank remains in the uppermost position by itself If not tighten the adjustment screw Move the arm into the park position Lower the unit to the floor and block the rear wheels Check the arm and single tank movement If necessary adjust according to sectio...

Page 87: ...ing wheels Jack up and block the unit securely 5 cm above the floor This must be done properly so that the unit does not tip over Disconnect all cables ground connection and K68 according to Fig 13 Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may still be charged even if the system is switched OFF and the mains cable is disconnected Life ...

Page 88: ...MOBILETT Plus Plus E SPR8 215 898 02 Page 12 of 28 Siemens Elema AB Rev 01 01 02 SPS UD Solna Sweden 10 12 Mechanical systems Fig 14 49 7 mm ...

Page 89: ...plate should be installed in the reverse order Reinstall the turn plate Reinstall the shafts Reinstall the bearing rings Reinstall the locking rings Connect the ground cable and K68 Reinstall the supporting wheels Reattach the side covers and the cover with the cassette compartment Set switch S10 in OFF position to charge the capacitor bank Reattach the system upper cover Connect the unit to mains...

Page 90: ...cure the arm with a rope Block the rear wheels and apply the parking brake Fig 15 Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may still be charged even if the system is switched OFF and the mains cable is disconnected Life threatening electric shock hazard exists See Chap Protective measures when working on the capacitor bank Equipment r...

Page 91: ...e cable ties see Page 9 5 Remove the collimator Disconnect all electrical connections to the single tank Remove the screws and nuts Fig 17 bolts placed in the ends of the axis of rotation Lift up the fork and remove the single tank Do not turn the fork around The new single tank has to be installed in the same position as the old one Replace the locking nuts with new nuts enclosed with the new sin...

Page 92: ...on Page 9 5 and 9 6 Tighten the adjustment screw Fig 16 so that the single tank is balanced Set switch S10 in the OFF position Fig 18 MOB00479 BK Black BU Blue RD Red YE Yellow GN YE Green yellow WH White VT Violet OG Orange BN Brown Voltage divider HT BK WH BK OG YE RD GN BK BK BK Temperature sensor HT transformer Resistor for mA measuring mA pos Stator for the rotation BK BU RD BN BK GN YE BK BN...

Page 93: ...cement of collimator Reattach all covers on the single tank Adjust the adjustment screw so that the single tank is balanced If any problems occur refer to the section Adjustment of arm system Use the service program P16 Show the exposure counter and record the values of the exposure counter in the service protocol and in the unit documentation Use the service program P10 Fast adaptation to adjust ...

Page 94: ...e support arm with a piece of rope Fig 19 Remove covers A and B Remove collimator cover C Fig 19 Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may still be charged even if the system is switched OFF and the mains cable is disconnected Life threatening electric shock hazard exists See Chap Protective measures when working on the capacitor b...

Page 95: ...ct all electrical connections to the single tank Remove the supporting hub in the rotation axis of the fork Remove all cable clamps and cable ties on the arm and the chassis Disconnect the single tank and the fork from the arm Disassemble the fork so the cables can be removed Disconnect the chassis connections and remove the cables Fig 20 Fig 20 Danger of injury The arm may move uncontrollably whe...

Page 96: ...Install the collimator and its wires Attach only cover C Connect all connections and reinstall all cable clamps Install new cable ties Fig 21 MOB00479 BK Black BU Blue RD Red YE Yellow GN YE Green yellow WH White VT Violet OG Orange BN Brown Voltage divider HT BK WH BK OG YE RD GN BK BK BK Temperature sensor HT transformer Resistor for mA measuring mA pos Stator for the rotation BK BU RD BN BK GN ...

Page 97: ...n the handle and the foot while keeping the single tank on the table Tighten the adjustment screw until the single tank and arm are balanced Refer to the section Adjustment of arm system Switch on the collimator lamp and check the compliance of the light and radiation fields according to the specifications in the section Replacement of collimator Reattach the remaining covers Do not stretch the ca...

Page 98: ...ed Fig 22 Remove the old lining by loosening screw pos 1 Fig 22 When installing the new lining make sure that the lining is facing the right way If the braking linings meet the requirements and the braking force is still low check the tension of the brake wire Remount the rear wheels Fig 22 Never work with the system open if the capacitor is charging or has just been charged The capacitor bank may...

Page 99: ... released pos 2 Fig 23 until required brake force is obtained After adjustments make sure the brake wire link is oriented according to pos 2 Fig 22 Lock its position using the nuts pos 3 Fig 22 Replacing the supporting wheels 10 Jack up the unit with a wood block measuring 5 x 5 cm Block the rear wheels so that they cannot turn Remove the locking rings pos 1 Fig 24 Remove the wheel shaft so that t...

Page 100: ... lining 10 Remove the joint covers Remove the locking nut Remove the screw and the disc springs Replace the horizontal friction lining Remount the screw and the the disc springs fig 27 Secure the screw with the locking nut Adjust the horizontal friction linings Refer to the section Adjustment of the friction linings Page 10 4 Reattach all covers Fig 25 Locking nut Disc springs Friction lining ...

Page 101: ... if the system is switched OFF and the mains cable is disconnected Life threatening electric shock hazard exists See Chap Protective measures when working on the capacitor bank Equipment required Large caliper rope ceiling support The ceiling suspension must be able to hold a weight of 100 kg since the entire weight of the arm will be on the ceiling support once the spring is removed If not the ar...

Page 102: ...ical systems Loosen the adjustment screw so that the arm is no longer balanced however the spring mechanism remains in position Fig 27 Do not allow the spring mechanism to fall out Remove the screw and washer from the upper counter bearing Fig 28 Remove the friction linings Fig 27 Fig 28 MOB00626 Adjustment screw ...

Page 103: ... upper counter bearing Tighten the screws Check that there is no braking action on the counter bearings Tighten the adjustment screw Make sure that the spring mechanism is under tension before releasing the ceiling support Attach the column cover and side covers and the cover with the cassette compartment Check parallelism refer to section Parallelism Adjust counterbalance refer to section Counter...

Page 104: ...MOBILETT Plus Plus E SPR8 215 898 02 Page 28 of 28 Siemens Elema AB Rev 01 01 02 SPS UD Solna Sweden 10 28 Mechanical systems This page intentionally left blank ...

Page 105: ...11 1 Siemens Elema AB SPR8 215 898 02 Page 1 of 2 MOBILETT Plus Plus E Solna Sweden Rev 01 01 02 SPS UD Changes to previous version 11 No changes initial publication ...

Page 106: ...MOBILETT Plus Plus E SPR8 215 898 02 Page 2 of 2 Siemens Elema AB Rev 01 01 02 SPS UD Solna Sweden 11 2 Changes to previous version This page intentionally left blank ...