Page 1: ...Patient Monitoring Hardware Upgrade Installation Guide I n t el l i V u e M P 4 0 5 0 IntelliVue Patient Monitor MP40 50 M8003 9301A M8003 9301A ...

Page 2: ...Part Number M8003 9301A ...

Page 3: ...1 1 Introduction 3 2 Checking your Tools and Kit Hardware 5 3 Installing J13 J30 7 4 Installing J40 11 5 Installing E12 13 6 Installing E25 21 7 Verification of Upgrade 25 ...

Page 4: ...2 ...

Page 5: ...Mining and Manufacturing Company 3M Part Number 8210 In addition to the table mat the person disassembling the monitor should either 1 Wear a conductive wrist strap M Part No 2046 attached to the mat by the 1 2 m 4 wire included with it or 2 Stand on a floor mat with leather soled shoes Before handling any circuit boards firmly touch the exposed metal on the case to equalize the ground potentials ...

Page 6: ... kits for the IntelliVue MP40 50 patient monitors M8027A J13 Serial Interface MIB ready 2 ports J14 Parallel Printer Interface J22 PS 2 Interface 2 ports J23 Interface for the Remote SpeedPoint and Alarm Device J30 Flexible Nurse Call Relay J40 Advanced Interface Wireless Network SVGA Video Nurse Call Relay E12 Add four internal module slots E25 Add Battery operation ...

Page 7: ...ginning check that your upgrade kit has the hardware listed next to the appropriate option in the table below Kit Part Number Description J13 M8081 67501 Serial Interface MIB ready board J14 M8082 67501 Parallel Printer Interface board J22 M8086 67501 PS 2 Interface board J23 M8086 67521 Interface board for Remote SpeedPoint and Alarm Device J30 M8087 67501 Flexible Nurse Call Relay board J40 M809...

Page 8: ...2 Checking your Tools and Kit Hardware Hardware List 6 ...

Page 9: ... J22 PS 2 Interface J30 Flexible Nurse Call Relay Installation Instructions Removing the Blank Covers CAUTION Make sure to unplug the AC power cord and to remove the batteries before removing blank covers or I O boards Failure to remove the power cord and the batteries could cause damage to the CMU as it is still under power 1 Remove the two screws securing each of the blank covers and remove them...

Page 10: ...emoval tool located inside the battery compartment door to pull off the blank covers or other I O boards if boards need to be switched Installing the I O Boards Insert the I O board into the slot according to the following table Board Removal Tool Slot 1 Slot2 Slot 3 ...

Page 11: ...snapped properly into place inside the battery compartment door after each use I O Board Possible Configurations Name Possible I O slots 1 2 3 4 5 6 7 8 9 19 Input Device Interface 02 03 02 02 02 02 Remote Device IInterface 02 03 03 02 02 02 Parallel Printer Interface 02 03 x 03 03 02 Nurse Call Relay 02 03 x x 03 x 03 03 02 MIB RS232 2x 02 03 x x x 03 x x 03 x 03 03 2nd MIB RS232 2x 02 03 x x x x...

Page 12: ...3 Installing J13 J30 Installation Instructions 10 ...

Page 13: ...tem Interface board for the MP40 50 monitors Installation Instructions CAUTION Make sure to unplug the AC power cord and to remove the batteries before removing blank covers or I O boards Failure to remove the power cord and the batteries could cause damage to the CMU as it is still under power ...

Page 14: ...n remove the board using the board removal tool located in the battery compartment 2 Insert the advanced System Interface board into the slot and secure it with the two screws NOTE Ensure that the board removal tool is snapped properly into place inside the battery compartment door after each use Board Removal Tool ...

Page 15: ...TION Make sure to unplug the AC power cord and to remove the batteries before removing blank covers I O boards and the System Interface Board Failure to remove the power cord and the batteries could cause damage to the CMU as it is still under power 1 Remove the two screws securing each of the blank covers I O boards and System Interface board and remove them ...

Page 16: ...compartment door to pull off the I O boards and the System Interface Board Separating the Front and Back Half of the monitor 1 Remove the four screws at the back of the monitor the lower ones first NOTE Hold on to the display when loosening the screws so the display does not fall off Board Removal Tool ...

Page 17: ...o the front because otherwise the flex panel adapter could be damaged Reassembly Note Make sure the metal sheet on the back of the display is inserted into the noses of the monitor rear housing when reassembling the two halves NOTE The pictures in some of the following steps already contain an inserted internal module slot rack ...

Page 18: ...crew the screws holding the flex panel adapter cable and pull off the flex cable carefully towards you Removing the Main Board 1 If a battery board is included always remove it before the main board by removing the four screws on the board Battery Board ...

Page 19: ...le simply pull directly on the cable The MSL power supply and speaker cables have different security locks and must be unlocked in order to be removed 3 Remove the nine screws pull the main board out of the rack connector and remove the board by moving it towards the bottom of the monitor MSL flat ribbon Power Supply Cable Speaker Cable Navigation Point Connector not connected in this picture ECG ...

Page 20: ...on Instructions 18 Removing the Blank Cover 1 Pull the back end of the cover towards you and at the same time release the two snaps on the inside with a screwdriver to release the back end of the cover Pull here In this direction ...

Page 21: ... 19 2 Release the three remaining snaps with a screwdriver and remove the blank cover completely 3 Insert the four slot rack into the monitor and secure it with two screws 4 Reassemble the monitor by performing the above steps in reverse order ...

Page 22: ...nd that the board is connected properly to the rack connector The screws must be reinserted in the order specified in the picture above Note that the screws marked 1a must only be inserted first in non battery versions of the monitor In battery versions the battery board is attached here with long screws Screw number 3 is attached to a white spacer which is not shown in this picture Press firmly o...

Page 23: ... and Back Half of the monitor CAUTION Make sure to unplug the AC power cord before removing blank covers or I O boards Failure to remove the power cord could cause damage to the CMU as it is still under power 1 Remove the four screws at the back of the monitor the lower ones first NOTE Hold on to the display when loosening the screws so the display does not fall off ...

Page 24: ...the front because otherwise the flex panel adapter could be damaged Reassembly Note Make sure the metal sheet on the back of the display is inserted into the noses of the monitor rear housing when reassembling the two halves 3 Attach the battery board to the main board with 4 screws Battery Board long screw long screw short screws ...

Page 25: ...the monitor by perfroming the above steps in reverse order Exchanging the battery door The battery door must be exchanged if the symbol is not printed on it The battery door with the symbol should look like this 1 Open the battery door 2 Using a screwdriver push the two pins inwards to release the battery door 3 Insert the new battery door and push the two pins outwards to secure it ...

Page 26: ...6 Installing E25 Installation Instructions 24 ...

Page 27: ...pened i e front and back half of the monitor have not been separated 1 Perform Power On Test block If an upgrade has been performed where the unit has been opened i e front and back half of the monitor have been separated 1 Perform Power On Test block 2 Perform Safety Test block For specific performance tests or documentation requirements see the IntelliVue MP40 MP50 service guide ...

Page 28: ...7 Verification of Upgrade 26 ...