HITACHI 4500-Il INSTRUCTIONS

 

The following instructions provide a guide to performing simple tasks on the microscope. Initial sections 
(A to D) describe how to insert a sample and obtain an image. Subsequent sections provide information 
on more COni~!CX operations such as changing the accelerating voltage or the detecton Sections K and 
L provide infonnation on what to do at the end of a session. 

 

Compared with the JEOL 840 SEM there are more variable parameters available to the operator. 
Optimum imaging conditions will vary from user to user depending on their samples. It is 
advisable that you spend some time attempting to determine which conditions are best for your 
sample.

 

 
When using this microscope please note the following rules: 

 

 

 

RULE 1: DON'T TOUCH A CONTROL IF YOU ARE NOT SURE OF THE 

 OUTCOME 

OF 

THAT 

ACTION. 

      

 

RULE 2: NEVER, EVER FORCE ANYTHING BEYOND FINGER STRENGTH

 

 

RULE 3: IF IN DOUBT; ASK FOR HELP 

 

 
A. START-UP
 
1.  Turn on the visual display by turning toggle switch 

DISPLAY POWER 

to 1. (These 

toggle switches must be pulled out slightly before switching - DO NOT FORCE THEM) 

2.  Turn the toggle switch SC 

AIRLOCK VALVE 

from 

CLOSED 

to 

AUTO 

(OPEN)- a 

few seconds later the green light on the HV off/ready display will come on. If it doesn't please 
contact EM staff for help. 

 
B. INSERT SAMPLE 
 

1.  Put specimen stub onto specimen holder> inc exact mechanism will vary depending upon 
the specimen type. EM Unit staff will advise the best method for your sample when you art 
being trained. The specimen holder and stub MUST be set so that so that thtt top of the 
specimen is level with or below tIze gauge. 
2.  Ensure toggle switch on SC 

AIRLOCK VALVE DISPLAY 

is set to S.E.C. 

3. Press 

AIR 

on 

Sc AIRLOCK VALVE DISPLAY - 

this will bleed air into the specimen 

exchange chamber. It will take 2~3O seconds for the chamber to come to air.

 

4.  Ensure stage settings 

(X,Y,Z,R 

and T) have been return to their default positions

 

(displayed on the specimen chamber). 
5.  Open specimen exchange chamber door Do not force. 
6.  Push specimen exchange rod forward slightly so it protrudes through the specimen exchange 
chamber door and screw specimen holder on to screw thread of the specimen exchange rod. 

PLEASE MAKE SURE YOU DO NOT GET 

CROSS THREAD. 

Pull specimen 

exchange rod back into specimen exchange chamber door. 

7.  Close airlock door 

Summary of Contents for McDATA Sphereon 4500

Page 1: ... switching DO NOT FORCE THEM 2 Turn the toggle switch SC AIRLOCK VALVE from CLOSED to AUTO OPEN a few seconds later the green light on the HV off ready display will come on If it doesn t please contact EM staff for help B INSERT SAMPLE 1 Put specimen stub onto specimen holder inc exact mechanism will vary depending upon the specimen type EM Unit staff will advise the best method for your sample wh...

Page 2: ...do this by manually turning the MAGNIFICATION knob or pressing the PRESET switch on the magnification display 8 Locate area of interest using X and V stage shifts you may need to go to low magnification see section I 9 Turn the COARSE FOCUS knob until a nominally focused image is obtained 10 Press ABC button on the image adjust section to automatically correct brightness and contrast 11 Press SMAL...

Page 3: ...ace sensitive information and reduce the size of the electron beam specimen interaction volume I Press FEll Vace 2 Using numeric keyboard key in desired accelerating voltage e g 5kV 3 Press ENTER 4 Microscope will automatically go to 5kV 5 Press PF16 EXIT to clear screen 6 Refocus image as described above 7 Go to ALIGN COLUMN section F F TO ALIGN THE COLUMN 1 Obtain a focused image at about 510 00...

Page 4: ...layed at INPUT will become active 5 Press PF16 EXIT to finish 6 Changing detectors may require you refocus the image and adjust brightness and contrast 7 n b The upper detector generally works better at short working distances 18 The lower detector generally works better at long working distances 18 See figure below You may accordingly need to change the specimen height H CHANGING THE WORKING DIST...

Page 5: ...urning the MAGNIFICATION knob or pressing the PRESET switch on the magnification display 3 Return X Y Z R and T controls to their preset levels listed on specimen chamber 4 Open airlock door by turning airlock level anticlockwise about one half turn DO NOT FORCE THIS LEVER it requires minimal effort to open 5 When airlock door is open fully insert specimen exchange rod into specimen chamber When r...

Page 6: ...ISHING A SESSION 1 Turn the SC AIRLOCK VALVE from AUTO OPEN to CLOSED the green light on the HV off ready display will turn off 2 Turn off the visual display by turning DISPLAY POWER to O 3 Fill in log book ...